Updated default run for intensity scans:
Results:
Link to analysis plan:
Beamtime run notes (Bhavna):
https://docs.google.com/document/d/1eYhR0q_LjSd4DGc9OnYuBl2aFFucNxmGOBq_E0NMT5g/edit
Analysis plan (Bhavna):
https://docs.google.com/spreadsheets/d/10m2vQLeCrlIA53khhRLaTRqm4YGobh0RGqGaE3A_WL8/edit
Primary finding for smalls: the detector is marginally timed in, which is causing the low values after switching. The delay, ACQToAsicR0delay, should be shifted from 18223 to 17223.
Google Drive folder for presentations with results etc.:
https://drive.google.com/drive/u/1/folders/1oolbkUqPAMD76qWZYiGQvVKIQPCqUfBS
Initial simulations of effects for getting the timing wrong. (Conny)
View file |
---|
name | Non-linearity when pulse arrives before the ACQ window.pptx |
---|
height | 250 |
---|
|
Overview of initial findings. (Conny)
View file |
---|
name | Start of analysis for mfxx1005021 – July 17.pptx |
---|
height | 250 |
---|
|
Register settings for the various cameras for easy access. (Conny)
View file |
---|
name | Register settings for differnet epix detectors.pptx |
---|
height | 250 |
---|
|
Small study of the double population. (Xavier)
View file |
---|
name | Double population Image and Timing.pptx |
---|
height | 250 |
---|
|
Timing in an ePix detector with LCLS timing system - VB (Dionisio)
View file |
---|
name | Timing in an ePix detector with LCLS timing system - VB.pptx |
---|
height | 250 |
---|
|
Double population (Xavier)
View file |
---|
name | Double population.pptx |
---|
height | 250 |
---|
|