Since the CCD operates optimally at -100C, it’s cooled down using a CryoTel Cooler with active vibration cancellation. The final piece of this system is an STA Archon controller which can receive configuration information from and send status and image data to a host PC via Ethernet connection.


Requirements set out by RIXS science staff:

    Parameter

Requirement

Format

4096 x 256, 1-d binnable

Frame rate

1Hz full-frame, 200Hz binned

Effective pixel size, dispersion

7um

Read noise

5 e- rms

Positional stability

100nm rms


Operating Manual for STA5200 Detector System