These are 2 identical detector systems commissioned from STA (STA5200 image sensor, SN31179 and SN31201) to enable qRIXS science in NEH 2.2. The CCD has 10um x 20um pixels, with 4224 pixels horizontally and 1024 pixels vertically, for a total image area of 42.24mm x 20.48mm. 

Since the CCD operates optimally at -100C, it’s cooled down using a CryoTel Cooler with active vibration cancellation. The final piece of this system is an STA Archon controller which can receive configuration information from and send status and image data to a host PC via Ethernet connection.


Requirements set out by RIXS science staff:

    Parameter

Requirement

Format

4096 x 256, 1-d binnable

Frame rate

1Hz full-frame, 200Hz binned

Effective pixel size, dispersion

7um

Read noise

5 e- rms

Positional stability

100nm rms


Operating Manual for STA5200 Detector System 

  • No labels