Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Nominal Operating Voltages and Currents

    
    

...

Environmental

Test conditions:

  • Detector running and stabilized for two hours, nominal settings.
  • Record thermometer and humidity sensor readings:

    Sensor   
    Airbox T39.64C  
    Airbox Humidity0.96%  
    Airbox Pressure1007 hPa  
    Dewpoint-25.4C  
    Chiller Flow2.3 l/m  
    Chiller Setpoint15C  

-        Electrical

  • Document grounding scheme
  • Read/record all available voltages and currents

Power Supply

 
Temp

29.3C

Voltage

11.792V

Current25.509A

 

-        Performance

  • Force high gain mode
  • Collect 1000 darks for analysis (Conny test 1; pedestals, noise, bad pixel map)
  • Force medium gain
  • Collect 1k darks (C2, pedestals/offsets, noise, bad pixel map)
  • Force low gain
  • Collect 1k darks (C3, pedestals/offsets, noise, bad pixel map)

...