Camera Model | Pulse Polarity | Pulse Delay | Pulse Width | comments |
---|---|---|---|---|
Adimec Opal 1000 | Pos | 0.0008 | 0.0002 | Pulse width affects exposure time. 0.0001 sec is too short. |
Adimec Opal 4000 | Pos | 0.0008 | 0.0002 | Maximum trigger rate is 30Hz. |
Pulnix TM-6740CL | Neg | 0.0006 | 0.0005 | Pulse width affects exposure time. |
LBL FCCD | Pos |
|
| Pulse width does not affect exposure time. |
Cspad | Neg | 2.5e-4 after Run Delay | 1e-5 | The integration window in the full cspad is positioned by the sum of |
Cspad2x2 | Pos | 5e-4 | 1e-5 | Run trigger the same, 5e-4 seconds, or 59524 ticks in the Daq system (typically evrstandalone process) Start with 11020 runTriggerDelay in the configuration (ONLY FOR SCANNING TIMING! |
Febo (normal) | Pos | 7.78e-4 | 1e-6 | For normal mode, trigger width not critical |
Febo (Average) | Pos | 4.41e-4 | 1e-6 | For averaging mode, trigger width not critical |
IPIMB | Neg | 0.000651824 | 5e-06 |
|
Zyla/Neo | Pos | 0.0007 | 1e-05 | For camera in 'overlap' mode using the global shutter only. Pulse width does not affect exposure time.
For details on other modes please see: Zyla Trigger Timing and Readout Mode Configuration |
Jungfrau | Pos | 0.00065 | 1e-05 | N.B. This works when using the standard internal delay of the Jungfrau of 0.000238 s |
...