L0M22 half module with pinholes pulled
High Voltage Test
(V) (uA)
V:D/I:D
50 0.43
100 0.67
200 0.89
300 1.10
400 1.58
475
current increased above 20uA when ramping from 400 to 500. Went down to 470, and current went to about 9 uA immediately, then decayed to 3 uA after about 1 minute wait. Able to ramp *slowly* (5V steps) to 500V. At 500V, current hit 20uA and then decayed to below 8 uA after about 1 min.
500 2.50 after 5 minutes
Low Voltage Test
(V) (A)
V:D/I:D
Pre-Config
AVDD: 2.50 0.106
DVDD: 2.50 0.245
V125: 1.25 0.000
Post-Config
AVDD: 2.50 0.262
DVDD: 2.50 0.245
V125: 1.25 0.230
Bad Channel Table
Sector B Sector A
pCH | apvCH | 128+apvCH | 2*128-apvCH-1| 128-apvCH-1
APV0
0 | 0 | 128 | 255 | 127
49 | 49 | 177 | 206 | 78
50 | 50 | 178 | 205 | 77
APV1
225 | 97 | 225 | 158 | 30
APV2
256 | 0 | 128 | 255 | 127
286 | 30 | 158 | 225 | 97
359 | 103 | 231 | 152 | 24
APV3
410 | 26 | 154 | 229 | 101
424 | 40 | 168 | 215 | 87
511 | 127 | 255 | 128 | 0
Nlow: 10
0V Bias:
50V Bias: