Search/Navigation:
Related:
SLAC/EPP/HPS Public
Jefferson Lab/Hall B/HPS Run Wiki
S30XL-LESA/LDMX
20190419
L0M21 half-module with pinholes pulled
High Voltage Test
(V) (uA)
V:D/I:D
50 0.50
100 0.70
125 0.67
150 2.21
175 9.08
200 18.74
LOW BREAKDOWN VOLTAGE!
Low Voltage Test
(V) (A)
V:D/I:D
Pre-Config
AVDD: 2.50 0.106
DVDD: 2.50 0.243
V125: 1.25 0.000
Post-Config
AVDD: 2.50 0.310
DVDD: 2.50 0.246
V125: 1.25 0.232
Bad Channel Table
Sector B Sector A
pCH | apvCH | 128+apvCH | 2*128-apvCH-1| 128-apvCH-1
APV0
0 | 0 | 128 | 255 | 127
65 | 65 | 193 | 190 | 62
APV1
171 | 43 | 171 | 212 | 84
179 | 51 | 179 | 204 | 76
180 | 52 | 180 | 203 | 75
181 | 53 | 181 | 202 | 74
182 | 54 | 182 | 201 | 73
183 | 55 | 183 | 200 | 72
184 | 56 | 184 | 199 | 71
210 | 82 | 210 | 173 | 45
215 | 87 | 215 | 168 | 40
APV2
257 | 1 | 129 | 254 | 126
260 | 4 | 132 | 251 | 123
269 | 13 | 141 | 242 | 114
278 | 22 | 150 | 233 | 105
279 | 23 | 151 | 232 | 104
295 | 39 | 167 | 216 | 88
368 | 112 | 240 | 143 | 15
APV3
389 | 5 | 133 | 250 | 122
469 | 85 | 213 | 170 | 42
487 | 103 | 231 | 152 | 24
511 | 127 | 255 | 128 | 0
Nlow: 22
0V Bias:
50V Bias: