Search/Navigation:
Related:
SLAC/EPP
/HPS Public
Jefferson Lab/Hall B
/HPS Run Wiki
S30XL-LESA/LDMX
Numbers are based on measured (from Test Run), nominal and maximum current draw estimates.
Max - 8.3W
Nom - 3.9W
Meas - 2.1W
*Ben Reese (10/8/2013)*As mentioned on the SVT DAQ meeting on Friday, I’d like to get some new worst case current estimates for DVDD, AVDD and V1_25 on the Hybrids.
The current numbers I’ve been designing towards are:
AVDD – 1.455A Max, 0.690 Nom, 0.420 Meas
DVDD – 1.455A Max, 0.690 Nom, 0.260 Meas
V1_25 – 0.825A Max, 0.325 Nom, 0.320 Meas
The max numbers seem rather high relative to the others.