Parameters
- Pixel size: 13.5um X 13.5um
- Sensor size: 27.6 mm X 27.6mm
- Gap size: 10.9 mm
- Array shape (2, 2048, 2048) or re-binned, ex. (2, 512, 512)
Data access
psana -m EventKeys -n 5 exp=sxrk4816:run=1
EventKey(type=Psana::Andor3d::FrameV1, src=DetInfo(SxrEndstation.0:DualAndor.0), alias="andorDual")
EventKey(type=Psana::Andor3d::ConfigV1, src=DetInfo(SxrEndstation.0:DualAndor.0), alias="andorDual")
# Get n-d array average:
det_ndarr_average -d exp=sxrk4816:run=1 -s SxrEndstation.0:DualAndor.0 -f nda-andor3d -p -v
Data content
Data with 2048x2048 bins exp=sxrk4816:run=1
source= Source("SxrEndstation.0:DualAndor.0")
shotIdStart = 18411
readoutTime = 1.91700208187
temperature = [ 8. 10.]
Configuration: <psana.Andor3d.ConfigV1 object at 0x7f087a9bee60>
width = 2048
height = 2048
numSensors = 2
orgX = 0
orgY = 0
binX = 1
binY = 1
exposureTime = 0.0010000000475
coolingTemp = 25.0
fanMode = ENUM_FAN_FULL
baselineClamp = 0
highCapacity = 0
gainIndex = 0
readoutSpeedIndex = 1
exposureEventCode = 80
exposureStartDelay = 12500
numDelayShots = 1
frameSize = 16777232
numPixelsX = 2048
numPixelsY = 2048
numPixelsPerSensor = 4194304
numPixels = 8388608
Data with 512x512 bins exp=sxrk4816:run=7
shotIdStart = 18366
readoutTime = 1.14359664917
temperature = [ 15. 13.]
Configuration: <psana.Andor3d.ConfigV1 object at 0x7fae382f8e60>
width = 2048
height = 2048
numSensors = 2
orgX = 0
orgY = 0
binX = 4
binY = 4
exposureTime = 0.541666686535
coolingTemp = 25.0
fanMode = ENUM_FAN_FULL
baselineClamp = 0
highCapacity = 0
gainIndex = 0
readoutSpeedIndex = 1
exposureEventCode = 80
exposureStartDelay = 12000
numDelayShots = 60
frameSize = 1048592
numPixelsX = 512
numPixelsY = 512
numPixelsPerSensor = 262144
numPixels = 524288
Geometry file example
/reg/g/psdm/detector/alignment/andor3d/calib-andor3d-2016-02-09/calib/Andor3d::CalibV1/SxrEndstation.0:DualAndor.0/geometry/7-end.data
# TITLE Geometry parameters of Andor3d
# DATE_TIME 2016-02-09 09:00:00 PDT
# METROLOGY no metrology available
# AUTHOR dubrovin
# EXPERIMENT SXR
# DETECTOR SxrEndstation.0:DualAndor.0
# CALIB_TYPE geometry
# COMMENT:01 This is a first file for Andor3d made by hands
# COMMENT:02 All translation and rotation pars of the object are defined w.r.t. parent object Cartesian frame
# COMMENT:03 Detector geometry: Sensor size 27.6 x 27.6mm, Pixel size 13.5 x 13.5um, distance between chips is 10.9mm.
# PARAM:01 PARENT - name and version of the parent object
# PARAM:02 PARENT_IND - index of the parent object
# PARAM:03 OBJECT - name and version of the object
# PARAM:04 OBJECT_IND - index of the new object
# PARAM:05 X0 - x-coordinate [um] of the object origin in the parent frame
# PARAM:06 Y0 - y-coordinate [um] of the object origin in the parent frame
# PARAM:07 Z0 - z-coordinate [um] of the object origin in the parent frame
# PARAM:08 ROT_Z - object design rotation angle [deg] around Z axis of the parent frame
# PARAM:09 ROT_Y - object design rotation angle [deg] around Y axis of the parent frame
# PARAM:10 ROT_X - object design rotation angle [deg] around X axis of the parent frame
# PARAM:11 TILT_Z - object tilt angle [deg] around Z axis of the parent frame
# PARAM:12 TILT_Y - object tilt angle [deg] around Y axis of the parent frame
# PARAM:13 TILT_X - object tilt angle [deg] around X axis of the parent frame
# HDR PARENT IND OBJECT IND X0[um] Y0[um] Z0[um] ROT-Z ROT-Y ROT-X TILT-Z TILT-Y TILT-X
ANDORDET:V1 0 MTRX:512:512:54:54 0 5400 0 0 90 0 0 0.0 0.0 0.0
ANDORDET:V1 0 MTRX:512:512:54:54 1 -33000 0 0 90 0 0 0.0 0.0 0.0
IP:V1 0 ANDORDET:V1 0 0 27600 0 -90 0 0 0.0 0.0 0.0
In order to support geometry of matrix-like sensors, the PSCalib.SegGeometryMatrixV1 object in PSCalib.SegGeometryStore.Create is initialized using variable shape sensor parameters coded in its name, for example MTRX:2048:2048:13.5:13.5, where values separated by colons stand for number of rows, columns, pixel size in rows and in columns in [um].
Images (Detector.image()) generated for geometry (2, 2048, 2048) and (2, 512, 512) using MTRX:2048:2048:13.5:13.5 or MTRX:512:512:54:54, respectively
![](/download/attachments/206768404/image-sxrk4816-r1-andor3d.png?version=1&modificationDate=1455053531000&api=v2)
![](/download/attachments/206768404/img-sxrk4816-r0007-Andor3d-512x512.png?version=1&modificationDate=1456192158000&api=v2)
References