Content
Optical measurement
- 2014-03-19-CSPAD-CXI-DS1-Metrology.xlsx
- 2014-03-19-CSPAD-CXI-DS1-Metrology.txt
- 2014-03-19-CSPAD-CXI-DS1-Metrology-corr.txt - for Q0 and Q2 use the same corrections of typos like in 2013-12-20 CSPAD DS1 Alignment
Quality check table for corrected Q0 and Q2 only
pair: S1 S2 dS1 dS2 L1 L2 dL1 dL2 <dS/L> angle(deg) D1 D2 dD d(dS) d(dL) dz3(um) Quad 0 pair: 0 20913 20911 35 33 43542 43545 -17 -14 0.00078 0.04474 48305 48304 1 2 -3 -15.007 pair: 1 20907 20911 6 10 43544 43547 -6 -3 0.00018 0.01053 48304 48305 -1 -4 -3 -9.007 pair: 2 20907 20908 -12 -13 43543 43545 -4 -2 -0.00029 -0.01645 48305 48300 5 1 -2 -20.006 pair: 3 20914 20915 -21 -22 43552 43561 -19 -10 -0.00049 -0.02828 48313 48321 -8 1 -9 -13.023 pair: 4 20909 20909 -39 -39 43597 43545 23 -29 -0.00090 -0.05128 48308 48347 -39 0 52 151.804 pair: 5 20904 20898 -24 -30 43557 43544 23 10 -0.00062 -0.03552 48309 48303 6 6 13 57.012 pair: 6 20934 20934 -256 -256 43578 43580 -127 -125 -0.00587 -0.33658 48344 48349 -5 0 -2 -3.001 pair: 7 20912 20942 -33 -63 43550 43548 -23 -25 -0.00110 -0.06315 48315 48317 -2 30 2 -44.030 Quad 1 pair: 0 20906 20905 68 67 43535 43541 -34 -28 0.00155 0.08883 48298 48295 3 1 -6 -0.000 pair: 1 20904 20902 -10 -12 43543 43537 6 0 -0.00011 -0.00658 48295 48299 -4 2 6 -1.013 pair: 2 20907 20900 17 24 43538 43537 11 10 0.00047 0.02698 48296 48295 1 -7 1 0.993 pair: 3 20902 20903 31 30 43539 43537 15 13 0.00070 0.04014 48295 48296 -1 1 2 -9.001 pair: 4 20906 20907 -1 0 43545 43545 -1 -1 -0.00001 -0.00066 48302 48304 -2 -1 0 -0.999 pair: 5 20909 20909 52 52 43542 43542 -27 -27 0.00119 0.06843 48300 48303 -3 0 0 -27.000 pair: 6 20907 20907 45 45 43542 43577 -14 21 0.00103 0.05919 48300 48333 -33 0 -35 7.075 pair: 7 20908 20906 57 59 43541 43543 27 29 0.00133 0.07632 48301 48301 0 -2 -2 -2.996 Quad 2 pair: 0 20909 20907 -29 -31 43543 43541 14 12 -0.00069 -0.03948 48300 48302 -2 2 2 50.998 pair: 1 20921 20912 -177 -186 43548 43563 87 102 -0.00417 -0.23876 48324 48311 13 9 -15 27.997 pair: 2 20972 20972 -81 -81 43524 43560 -18 18 -0.00186 -0.10659 48364 48294 70 0 -36 -50.086 pair: 3 20907 20907 -51 -51 43542 43543 -23 -22 -0.00117 -0.06711 48303 48299 4 0 -1 4.999 pair: 4 20908 20905 127 124 43542 43541 -60 -61 0.00288 0.16514 48300 48300 0 3 1 -59.003 pair: 5 20903 20931 76 104 43546 43545 -42 -43 0.00207 0.11842 48309 48308 1 -28 1 20.900 pair: 6 20911 20910 74 75 43547 43546 37 36 0.00171 0.09802 48307 48306 1 -1 1 9.000 pair: 7 20918 20925 149 142 43551 43540 80 69 0.00334 0.19144 48315 48306 9 7 11 -30.015 Quad 3 pair: 0 20902 20899 72 69 43537 43538 -33 -32 0.00162 0.09278 48295 48293 2 3 -1 -5.015 pair: 1 20905 20909 48 52 43534 43545 -28 -17 0.00115 0.06580 48300 48297 3 -4 -11 17.022 pair: 2 20907 20907 259 259 43539 43539 124 124 0.00595 0.34083 48299 48299 0 0 0 -4.000 pair: 3 20894 20901 16 9 43538 43537 10 9 0.00029 0.01645 48296 48289 7 7 1 -4.004 pair: 4 20904 20903 95 94 43541 43542 -46 -45 0.00217 0.12435 48299 48299 0 1 -1 -17.002 pair: 5 20904 20905 158 159 43541 43542 -77 -76 0.00364 0.20857 48299 48300 -1 -1 -1 1.001 pair: 6 20903 20905 41 39 43540 43539 22 21 0.00092 0.05264 48299 48295 4 2 1 -12.006 pair: 7 20907 20904 11 14 43542 43543 5 6 0.00029 0.01645 48300 48301 -1 -3 -1 -25.995
Quality for re-measured Quads Q1 and Q3 is very good
Geometry parameters
center/0-end.data (pixel)
197.94 198.05 304.45 97.98 623.08 623.74 706.30 498.64 197.93 198.08 304.13 96.70 621.81 621.77 704.64 497.90 197.22 198.51 300.71 96.77 621.46 620.95 709.39 496.88 197.69 197.92 301.97 99.21 623.82 622.78 704.38 497.16 301.22 95.14 621.13 621.36 516.12 724.42 197.91 196.12 303.22 95.04 620.07 620.30 518.10 725.83 198.43 198.14 301.51 94.34 622.96 623.24 519.67 725.08 198.86 198.69 301.11 95.32 622.75 622.42 517.86 724.77 197.25 197.98 0.47 0.17 -0.25 0.17 0.65 -0.19 -0.14 -0.50 1.62 0.56 2.34 2.25 1.47 2.11 1.02 0.97 0.01 -0.37 0.95 -0.22 0.08 -0.75 0.21 0.05 0.93 0.37 2.47 2.46 1.77 2.24 0.55 1.07
or the same in um
21757 21769 33464 10769 68489 68561 77637 54810 21757 21773 33430 10628 68349 68345 77454 54729 21678 21820 33054 10637 68310 68254 77976 54617 21730 21755 33193 10904 68570 68456 77425 54648 33110 10457 68275 68299 56732 79628 21754 21558 33329 10446 68158 68183 56949 79783 21811 21779 33141 10369 68476 68506 57121 79700 21858 21839 33098 10477 68452 68416 56923 79666 21681 21761 51 18 -28 18 71 -20 -15 -54 178 61 257 247 161 231 111 106 1 -40 104 -24 9 -82 23 6 102 40 272 270 194 246 60 118
tilt/0-end.data (degree)
0.04474 0.01053 -0.01645 -0.02828 -0.05128 -0.03552 -0.33658 -0.06315 0.08883 -0.00658 0.02698 0.04014 -0.00066 0.06843 0.05919 0.07632 -0.03948 -0.23876 -0.10659 -0.06711 0.16514 0.11842 0.09802 0.19144 0.09278 0.06580 0.34083 0.01645 0.12435 0.20857 0.05264 0.01645
2014-03-22 Tuning of quads
use exp=cxii0114:run=227
offset_corr/0-end.data
0 1 2 5 0 2 5 4 0 0 0 0
2014-03-24 Tuning of quads
use exp=cxii0114:run=227
At more detailed consideration with semi-automatic ring-center finding algorithm (and masking quads etc.) it was revealed that the gap between quads need to be increased by a few pixels, that is set in
offset_corr/0-end.data
0 0 9 12 0 5 10 3 0 0 0 0
which gives image, r-phi plot, and radial histogram:
Overview
Content Tools