Content

Optical measurement

Plots from Gabriel Blaj:

 

Quality check

Evaluated parameters

We calculate
S1 - 1st short side length of 2x1
S2 - 2nd short side length of 2x1
L1 - 1st long side length of 2x1
L2 - 2nd long side length of 2x1
D1 - 1st diagonal of 2x1 between corners 1 and 3
D2 - 2nd diagonal of 2x1 between corners 2 and 4
dS and dL are the deviations of the 1st and 2nd corner along the short and long sides, respectively. The sign of all dS are chosen in order to provide correct sign for the tilt angle (the same direction for all 2x1 sensors).
<dS/L> - the tilt angle of 2x1 averaged over two sides in radians.
angle(deg) - the same angle in degrees.
dD = D1 - D2
d(dS) = dS1 - dS2
d(dL) = dL1 - dL2
dz3(um) - signed distance from 2x1 sensor plane and corner 3, where the 2x1 sensor plane contains the corner points p1, p2, and p4. This plane is defined by the vectors v21=p2-p1, v41=p4-p1, and their orthogonal vector

vort = [v21 x v41].

Scalar product with normalization defines the distance from point 3 to the 2x1 plane containing 3 other points:

dz3 = (v31 * vort) / |vort|.

S1=S2, L1=L2 - the 2x1 sides should have equal length and width,
D1=D2 - the 2x1 diagonals should be equal,
dS1 = dS2 (approximately)= (388/185)*dL1 = (388/185)*dL2 - tilt angle should provide consistent deviation for all corners,
dD=0, d(dS)=0, and d(dL)=0 - within precision of measurement.
dz3(um) = 0

Everything, excluding <dS/L> and angle(deg), are in micrometers.

 

Quality check table for original metrology file

pair:        S1      S2     dS1     dS2        L1      L2     dL1     dL2    <dS/L>  angle(deg)      D1      D2      dD   d(dS)   d(dL)    dz3(um)

Quad  0
pair: 0   20913   20911      35      33     43542   43545     -17     -14    0.00078    0.04474   48305   48304       1       2      -3    -15.007
pair: 1   20907   20911       6      10     43544   43547      -6      -3    0.00018    0.01053   48304   48305      -1      -4      -3     -9.007
pair: 2   20907   20908     -12     -13     43543   43545      -4      -2   -0.00029   -0.01645   48305   48300       5       1      -2    -20.006
pair: 3   20914   20915     -21     -22     43552   43561     -19     -10   -0.00049   -0.02828   48313   48321      -8       1      -9    -13.023
pair: 4   21517   20909     569     -39     43597   43545      23     -29    0.00608    0.34824   48574   48347     227     608      52    149.012
pair: 5   20904   20898     -24     -30     43557   43544      23      10   -0.00062   -0.03552   48309   48303       6       6      13     57.012
pair: 6   21711   20934    -256     521     43578   43580    -127    -125    0.00304    0.17419   48344   48687    -343    -777      -2     -3.071
pair: 7   20912   20942     -33     -63     43550   43548     -23     -25   -0.00110   -0.06315   48315   48317      -2      30       2    -44.030

Quad  1
pair: 0   20906   20915      24      33     43541   43541     -13     -13    0.00065    0.03750   48302   48301       1      -9       0     17.006
pair: 1   20908   20915     -52     -45     43544   43540      24      20   -0.00111   -0.06382   48302   48304      -2      -7       4     -5.998
pair: 2   20850   20913     -15     -78     43544   43543      -6      -7   -0.00107   -0.06119   48305   48277      28      63       1     -6.022
pair: 3   20916   20909      -5       2     43543   43543      -2      -2   -0.00003   -0.00197   48303   48305      -2      -7       0    -14.003
pair: 4   20911   20864     -35     -82     43543   43545      40      42   -0.00134   -0.07697   48306   48282      24      47      -2     17.042
pair: 5   20911   20900      16       5     43546   43605     -10      49    0.00024    0.01381   48352   48308      44      11     -59    -63.016
pair: 6   20906   20913     120     113     43550   43550      53      53    0.00268    0.15327   48307   48312      -5       7       0      8.996
pair: 7   20909   20909      42      42     43515   43548     -13      20    0.00096    0.05528   48277   48307     -30       0     -33     12.977

Quad  2
pair: 0   20909   20907     -29     -31     43543   43541      14      12   -0.00069   -0.03948   48300   48302      -2       2       2     50.998
pair: 1   20921   20912    -177    -186     43548   43563      87     102   -0.00417   -0.23876   48324   48311      13       9     -15     27.997
pair: 2   20972   21502     -81    -611     43524   43560     -18      18   -0.00794   -0.45515   48597   48294     303     530     -36    -51.477
pair: 3   20907   20907     -51     -51     43542   43543     -23     -22   -0.00117   -0.06711   48303   48299       4       0      -1      4.999
pair: 4   20908   20905     127     124     43542   43541     -60     -61    0.00288    0.16514   48300   48300       0       3       1    -59.003
pair: 5   20903   20931      76     104     43546   43545     -42     -43    0.00207    0.11842   48309   48308       1     -28       1     20.900
pair: 6   20911   20910      74      75     43547   43546      37      36    0.00171    0.09802   48307   48306       1      -1       1      9.000
pair: 7   20918   20925     149     142     43551   43540      80      69    0.00334    0.19144   48315   48306       9       7      11    -30.015

Quad  3
pair: 0   20908   20908     -83     -83     43541   43542      38      39   -0.00191   -0.10922   48300   48302      -2       0      -1    130.998
pair: 1   20908   20908    -104    -104     43545   43545      50      50   -0.00239   -0.13684   48304   48304       0       0       0      9.000
pair: 2   20913   20912     103     104     43550   43546      53      49    0.00238    0.13617   48310   48307       3      -1       4      2.000
pair: 3   20734   20915    -131    -312     43547   43565     -84     -66   -0.00508   -0.29135   48306   48250      56     181     -18     -9.111
pair: 4   20909   20908     -57     -58     43547   43612      30      95   -0.00132   -0.07560   48367   48304      63       1     -65     -9.075
pair: 5   20914   20940     -15      11     43576   43549      21      -6   -0.00005   -0.00263   48334   48322      12     -26      27    -12.026
pair: 6   20905   20910     210     205     43544   43580      69     105    0.00476    0.27292   48308   48331     -23       5     -36     39.953
pair: 7   20918   20914     186     190     43565   43545      91      71    0.00432    0.24731   48308   48325     -17      -4      20      3.996

Significant deviations are observed in 4 points.

Applied corrections

  • Quad 0, pair 4, point 17 Y: 45689 + 608 = 46297
  • Quad 0, pair 6, point 32 X: 88753 - 777 = 87976
  • Quad 2, pair 2, point 15 X: 44111 - 530 = 43581
  • Quad 3, pair 3, point 12 X:21361 + 181 = 21542

 

Quality check table for corrected metrology file

pair:        S1      S2     dS1     dS2        L1      L2     dL1     dL2    <dS/L>  angle(deg)      D1      D2      dD   d(dS)   d(dL)    dz3(um)

Quad  0
pair: 0   20913   20911      35      33     43542   43545     -17     -14    0.00078    0.04474   48305   48304       1       2      -3    -15.007
pair: 1   20907   20911       6      10     43544   43547      -6      -3    0.00018    0.01053   48304   48305      -1      -4      -3     -9.007
pair: 2   20907   20908     -12     -13     43543   43545      -4      -2   -0.00029   -0.01645   48305   48300       5       1      -2    -20.006
pair: 3   20914   20915     -21     -22     43552   43561     -19     -10   -0.00049   -0.02828   48313   48321      -8       1      -9    -13.023
pair: 4   20909   20909     -39     -39     43597   43545      23     -29   -0.00090   -0.05128   48308   48347     -39       0      52    151.804
pair: 5   20904   20898     -24     -30     43557   43544      23      10   -0.00062   -0.03552   48309   48303       6       6      13     57.012
pair: 6   20934   20934    -256    -256     43578   43580    -127    -125   -0.00587   -0.33658   48344   48349      -5       0      -2     -3.001
pair: 7   20912   20942     -33     -63     43550   43548     -23     -25   -0.00110   -0.06315   48315   48317      -2      30       2    -44.030

Quad  1
pair: 0   20906   20915      24      33     43541   43541     -13     -13    0.00065    0.03750   48302   48301       1      -9       0     17.006
pair: 1   20908   20915     -52     -45     43544   43540      24      20   -0.00111   -0.06382   48302   48304      -2      -7       4     -5.998
pair: 2   20850   20913     -15     -78     43544   43543      -6      -7   -0.00107   -0.06119   48305   48277      28      63       1     -6.022
pair: 3   20916   20909      -5       2     43543   43543      -2      -2   -0.00003   -0.00197   48303   48305      -2      -7       0    -14.003
pair: 4   20911   20864     -35     -82     43543   43545      40      42   -0.00134   -0.07697   48306   48282      24      47      -2     17.042
pair: 5   20911   20900      16       5     43546   43605     -10      49    0.00024    0.01381   48352   48308      44      11     -59    -63.016
pair: 6   20906   20913     120     113     43550   43550      53      53    0.00268    0.15327   48307   48312      -5       7       0      8.996
pair: 7   20909   20909      42      42     43515   43548     -13      20    0.00096    0.05528   48277   48307     -30       0     -33     12.977

Quad  2
pair: 0   20909   20907     -29     -31     43543   43541      14      12   -0.00069   -0.03948   48300   48302      -2       2       2     50.998
pair: 1   20921   20912    -177    -186     43548   43563      87     102   -0.00417   -0.23876   48324   48311      13       9     -15     27.997
pair: 2   20972   20972     -81     -81     43524   43560     -18      18   -0.00186   -0.10659   48364   48294      70       0     -36    -50.086
pair: 3   20907   20907     -51     -51     43542   43543     -23     -22   -0.00117   -0.06711   48303   48299       4       0      -1      4.999
pair: 4   20908   20905     127     124     43542   43541     -60     -61    0.00288    0.16514   48300   48300       0       3       1    -59.003
pair: 5   20903   20931      76     104     43546   43545     -42     -43    0.00207    0.11842   48309   48308       1     -28       1     20.900
pair: 6   20911   20910      74      75     43547   43546      37      36    0.00171    0.09802   48307   48306       1      -1       1      9.000
pair: 7   20918   20925     149     142     43551   43540      80      69    0.00334    0.19144   48315   48306       9       7      11    -30.015

Quad  3
pair: 0   20908   20908     -83     -83     43541   43542      38      39   -0.00191   -0.10922   48300   48302      -2       0      -1    130.998
pair: 1   20908   20908    -104    -104     43545   43545      50      50   -0.00239   -0.13684   48304   48304       0       0       0      9.000
pair: 2   20913   20912     103     104     43550   43546      53      49    0.00238    0.13617   48310   48307       3      -1       4      2.000
pair: 3   20915   20915    -131    -131     43547   43565     -84     -66   -0.00301   -0.17232   48306   48328     -22       0     -18     -9.014
pair: 4   20909   20908     -57     -58     43547   43612      30      95   -0.00132   -0.07560   48367   48304      63       1     -65     -9.075
pair: 5   20914   20940     -15      11     43576   43549      21      -6   -0.00005   -0.00263   48334   48322      12     -26      27    -12.026
pair: 6   20905   20910     210     205     43544   43580      69     105    0.00476    0.27292   48308   48331     -23       5     -36     39.953
pair: 7   20918   20914     186     190     43565   43545      91      71    0.00432    0.24731   48308   48325     -17      -4      20      3.996
  • X-Y precision of measurements is about 60um
  • Tilt angles are in range from -0.33 to 0.27 degree
  • Z deviations from flatness are in the range from -63 to 152um

Geometry parameters

(tick) center/0-end.data (pixel)

 197.94   198.05   304.45    97.98   623.08   623.74   706.30   498.64  
 198.22   198.17   304.58    97.25   622.90   623.09   708.37   495.18  
 197.22   198.51   300.71    96.77   621.46   620.95   709.39   496.88  
 198.80   198.29   304.26   101.44   625.93   625.40   705.81   497.16  

 301.22    95.14   621.13   621.36   516.12   724.42   197.91   196.12  
 303.12    94.88   619.87   620.29   515.50   722.83   197.13   196.76  
 301.51    94.34   622.96   623.24   519.67   725.08   198.86   198.69  
 300.50    94.63   621.71   622.07   515.69   722.61   196.56   195.96  

   0.47     0.17    -0.25     0.17     0.65    -0.19    -0.14    -0.50  
   0.36    -0.16    -0.07     0.03     0.38    -0.13    -0.04    -0.18  
   0.01    -0.37     0.95    -0.22     0.08    -0.75     0.21     0.05  
  -0.05    -0.08     0.25     0.19     0.03    -0.21     0.21     0.12 

or the same in um

 21757   21769   33464   10769   68489   68561   77637   54810  
 21788   21783   33479   10689   68469   68489   77863   54430  
 21678   21820   33054   10637   68310   68254   77976   54617  
 21851   21796   33444   11150   68801   68743   77582   54648  

 33110   10457   68275   68299   56732   79628   21754   21558  
 33319   10429   68135   68182   56664   79453   21668   21628  
 33141   10369   68476   68506   57121   79700   21858   21839  
 33030   10402   68338   68378   56684   79429   21605   21540  

    51      18     -28      18      71     -20     -15     -54  
    39     -18      -7       3      41     -14      -4     -20  
     1     -40     104     -24       9     -82      23       6  
    -5      -8      27      20       2     -23      23      13

 

(tick) tilt/0-end.data (degree)

 0.04474   0.01053  -0.01645  -0.02828  -0.05128  -0.03552  -0.33658  -0.06315  
 0.03750  -0.06382  -0.06119  -0.00197  -0.07697   0.01381   0.15327   0.05528  
-0.03948  -0.23876  -0.10659  -0.06711   0.16514   0.11842   0.09802   0.19144  
-0.10922  -0.13684   0.13617  -0.17232  -0.07560  -0.00263   0.27292   0.24731 

 

(minus) offset/0-end.data

0    0  820  820
0  820  820    0
0    0    0    0

 

2014-03-05 Alignment of Quads

Data for alignment

In order to get cspad geometry parameters for experiment cxib6714 Jason Koglin suggested to use cxi83714 runs 17,18, 127, 128, 130-132, 135, 136.

CSPAD array with subtracted pedestal and threshold 4*rms(per pixel) is averaged over 5000 events of run 136.

Parameters from latest optical measurement

center and tilt are used from /reg/neh/home1/dubrovin/LCLS/CSPadAlignment-v01/calib-cxi-ds1-2013-12-20

 

Tuned calibration parameters

offset_corr/0-end.data :

  0   0    4   4
  0   4    1   1
  0   0    0   0

 

marg_gap_shift/0-end.data :

15   40   4  40
15   40   4  40
 0    0   0   0

Results

 

2014-03-26 quads alignment with semi-automatic algorithm for ring center finding

R-phi presentation if the image using previous alignment

shows that manual quads alignment is done with precision about ±1 pixel but can be improved.

Slightly better results for quads alignment were obtained with semi-automatic ring center finder by changing parameters

offset_corr/136-136.data :

  1   1    4   4
  0   5    2   2
  0   0    0   0

 

  • No labels