Content
Optical measurements
E-mail from Gabriel:
Attached, the new metrology file for the (refurbished) MEC quad delivered today. It should be applied to all experiments from today (June 8th, 2015) to the next time the camera will be disassembled.
As always, please use only Mikhail's curated geometry and not the raw geometry attached to this message.
Gabriel
- 2015-06-08-Quad1-MEC-Metrology.xlsx - file with optical measurements from Gabriel
- 2015-06-08-Quad1-MEC-Metrology.txt - text version of the same file
Quality check
Quality check in X-Y plane
2015-06-09 10:11:54 (info) CalibManager.GUIMetrology: Quality check in X-Y plane: segm: S1 S2 dS1 dS2 L1 L2 dL1 dL2 angle(deg) D1 D2 dD d(dS) d(dL) Quad 0 segm: 0 -20911 -20909 -92 -90 43543 43578 -49 -14 -0.11969 48330 48308 22 -2 -35 segm: 1 -20910 -20910 -45 -45 43549 43552 -22 -19 -0.05920 48311 48309 2 0 -3 segm: 2 -20909 -20905 -23 -19 -43541 -43568 7 -20 -0.02763 48327 48297 30 -4 27 segm: 3 -20905 -20908 29 26 -43540 -43545 -15 -20 0.03619 48305 48297 8 3 5 segm: 4 20946 20946 146 146 -43579 -43579 37 37 -0.19195 48381 48321 60 0 0 segm: 5 20909 20910 -163 -162 -43545 -43547 -77 -79 0.21381 48306 48306 0 -1 2 segm: 6 -20924 -20900 -13 11 -43547 -43585 17 -21 -0.00132 48327 48322 5 -24 38 segm: 7 -20914 -20901 63 76 -43540 -43546 -37 -43 0.09145 48308 48296 12 -13 6
Quality check in Z
2015-06-09 10:11:54 (info) CalibManager.GUIMetrology: Quality check in Z: segm: SA LA XSize YSize dZS1 dZS2 dZL1 dZL2 dZSA dZLA ddZS ddZL dZX dZY angXZ(deg) angYZ(deg) dz3(um) Quad 0 segm: 0 -20910 43560 43560 20910 -107 14 -81 40 -46 -20 -121 -121 -20 -46 -0.02631 -0.12605 120.956 segm: 1 -20910 43550 43550 20910 126 62 63 -1 94 31 64 64 31 94 0.04078 0.25757 -63.999 segm: 2 -20907 -43554 20907 43554 2 17 35 50 9 42 -15 -15 9 42 0.02466 0.05525 -14.972 segm: 3 -20906 -43542 20906 43542 -17 -8 43 52 -12 47 -9 -9 -12 47 -0.03289 0.06185 -8.995 segm: 4 20946 -43579 43579 20946 -28 -14 -37 -23 -21 -30 -14 -14 -30 -21 -0.03944 -0.05744 14.000 segm: 5 20909 -43546 43546 20909 -37 -38 -28 -29 -37 -28 1 1 -28 -37 -0.03684 -0.10139 -0.997 segm: 6 -20912 -43566 20912 43566 -4 26 11 41 11 26 -30 -30 11 26 0.03014 0.03419 -29.994 segm: 7 -20907 -43543 20907 43543 -83 -77 29 35 -80 32 -6 -6 -80 32 -0.21924 0.04211 -5.947
Corrections
- No correction applied.
Constants
center
194.53 193.87 303.83 91.27 621.30 621.00 706.17 493.30 194.53 193.87 303.83 91.27 621.30 621.00 706.17 493.30 194.53 193.87 303.83 91.27 621.30 621.00 706.17 493.30 194.53 193.87 303.83 91.27 621.30 621.00 706.17 493.30 305.06 93.15 623.16 623.98 512.12 724.83 197.79 197.57 305.06 93.15 623.16 623.98 512.12 724.83 197.79 197.57 305.06 93.15 623.16 623.98 512.12 724.83 197.79 197.57 305.06 93.15 623.16 623.98 512.12 724.83 197.79 197.57 0.39 0.11 0.03 -0.03 0.15 0.11 0.44 -0.50 0.39 0.11 0.03 -0.03 0.15 0.11 0.44 -0.50 0.39 0.11 0.03 -0.03 0.15 0.11 0.44 -0.50 0.39 0.11 0.03 -0.03 0.15 0.11 0.44 -0.50
tilt (degree)
-0.11969 -0.05920 -0.02763 0.03619 -0.19195 0.21381 -0.00132 0.09145 -0.11969 -0.05920 -0.02763 0.03619 -0.19195 0.21381 -0.00132 0.09145 -0.11969 -0.05920 -0.02763 0.03619 -0.19195 0.21381 -0.00132 0.09145 -0.11969 -0.05920 -0.02763 0.03619 -0.19195 0.21381 -0.00132 0.09145
geometry
# TITLE Geometry parameters of CSPAD # DATE_TIME 2015-06-09 10:11:54 PDT # METROLOGY /reg/g/psdm/detector/alignment/cspad/calib-mec-2015-06-08/work/2015-06-08-Quad1-MEC-Metrology.txt # AUTHOR dubrovin # EXPERIMENT MEC # DETECTOR CSPAD # CALIB_TYPE geometry # COMMENT:01 Table contains the list of geometry parameters for alignment of 2x1 sensors, quads, CSPAD, etc # COMMENT:02 All translation and rotation pars of the object are defined w.r.t. parent object Cartesian frame # PARAM:01 PARENT - name and version of the parent object # PARAM:02 PARENT_IND - index of the parent object # PARAM:03 OBJECT - name and version of the object # PARAM:04 OBJECT_IND - index of the new object # PARAM:05 X0 - x-coordinate [um] of the object origin in the parent frame # PARAM:06 Y0 - y-coordinate [um] of the object origin in the parent frame # PARAM:07 Z0 - z-coordinate [um] of the object origin in the parent frame # PARAM:08 ROT_Z - object design rotation angle [deg] around Z axis of the parent frame # PARAM:09 ROT_Y - object design rotation angle [deg] around Y axis of the parent frame # PARAM:10 ROT_X - object design rotation angle [deg] around X axis of the parent frame # PARAM:11 TILT_Z - object tilt angle [deg] around Z axis of the parent frame # PARAM:12 TILT_Y - object tilt angle [deg] around Y axis of the parent frame # PARAM:13 TILT_X - object tilt angle [deg] around X axis of the parent frame # HDR PARENT IND OBJECT IND X0[um] Y0[um] Z0[um] ROT-Z ROT-Y ROT-X TILT-Z TILT-Y TILT-X QUAD:V1 0 SENS2X1:V1 0 21382 33532 43 0 0 0 -0.11969 -0.02631 -0.12605 QUAD:V1 0 SENS2X1:V1 1 21310 10238 12 0 0 0 -0.05920 0.04078 0.25757 QUAD:V1 0 SENS2X1:V1 2 33397 68498 3 270 0 0 -0.02763 0.02466 0.05525 QUAD:V1 0 SENS2X1:V1 3 10032 68588 -3 270 0 0 0.03619 -0.03289 0.06185 QUAD:V1 0 SENS2X1:V1 4 68293 56292 16 180 0 0 -0.19195 -0.03944 -0.05744 QUAD:V1 0 SENS2X1:V1 5 68260 79673 12 180 0 0 0.21381 -0.03684 -0.10139 QUAD:V1 0 SENS2X1:V1 6 77622 21741 48 270 0 0 -0.00132 0.03014 0.03419 QUAD:V1 0 SENS2X1:V1 7 54223 21717 -55 270 0 0 0.09145 -0.21924 0.04211 ... CSPAD:V1 0 QUAD:V1 0 -4500 -4500 0 90 0 0 0.00000 0.00000 0.00000 CSPAD:V1 0 QUAD:V1 1 -4500 4500 0 0 0 0 0.00000 0.00000 0.00000 CSPAD:V1 0 QUAD:V1 2 4500 4500 0 270 0 0 0.00000 0.00000 0.00000 CSPAD:V1 0 QUAD:V1 3 4500 -4500 0 180 0 0 0.00000 0.00000 0.00000 #RAIL 0 CSPAD:V1 0 0 0 1000000 0 0 0 0.00000 0.00000 0.00000 #IP 0 RAIL 0 0 0 0 0 0 0 0.00000 0.00000 0.00000
Alignment check with rings
TBD
Deployment
Parameters for geometry
(and old style center
and tilt
) are available in
Archive: /reg/g/psdm/detector/alignment/cspad/calib-mec-2015-06-08/calib/CsPad::CalibV1/MecTargetChamber.0:Cspad.0/geometry/0-end.data
0-end.data -
non-corrected
Deployment: for all experiments from 2015-06-08 to 2015-...
Summary
All optic measurements are within 3σ of precision. No correction applied.
References
Overview
Content Tools