Identification Details and Hardware Changelog


Nominal Operating Voltages and Currents

Ch.

Line

V

I





Performance Measures

Experiment=detdaq18, Run=504 (Jan 28, 2021)

Pedestal Summary

Method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as xxxx.

Map of pedestal non-conforming pixels for all gain modes here

Gain Mode

Mean Pedestal Value (ADU)

Number of NCP

Pixels Map, Histogram

FH3066 ± 2061Graphs
FM3035 ± 2051Graphs
FL3021 ± 2041Graphs
AHL3070 ± 2061Graphs
AML3038 ± 205 1Graphs


Noise Summary

Map of noise non-conforming pixels for all gain modes here.  

Gain Mode

Variation (ADU)

Number of NCP

Pixels Map, Histogram

FH55533Graphs
FM21519Graphs
FL22705Graphs
AHL64276Graphs
AML32619Graphs


Gain and uniformity (photon source)

Map of gain non-conforming pixels for all gain modes here

Gain Mode

Number of NCP

FH0
FM0
FL480
AHL1
AML1

Linearity (signal injection)