Identification Details and Hardware Changelog


Nominal Operating Voltages and Currents

Ch.

Line

V

I





Performance Measures

Experiment=detdaq18, Run=427 (Nov 24, 2020)

Pedestal Summary

Method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as xxxx.

Map of pedestal non-conforming pixels for all gain modes here

Gain Mode

Mean Pedestal Value (ADU)

Number of NCP

Pixels Map, Histogram

FH2916 ± 170  3Graphs
FM2895 ± 169 3Graphs
FL2889 ± 1693Graphs
AHL2939 ± 1713Graphs
AML2914 ± 1703Graphs


Noise Summary

Map of noise non-conforming pixels for all gain modes here.  

Gain Mode

Variation (ADU)

Number of NCP

Pixels Map, Histogram

FH67206Graphs
FM354153Graphs
FL236520Graphs
AHL62146Graphs
AML310515Graphs


Gain and uniformity (photon source)

Map of gain non-conforming pixels for all gain modes here

Gain Mode

Number of NCP

FH0
FM0
FL481
AHL8
AML8

Linearity (signal injection)