Identification Details and Hardware Changelog

FirmwareCarrierID0CarrierID1DigitalCardID0DigitalCardID1AnalogCardID0AnalogCardID1Date
3925999620099643289735903242341232100352115453287526540884490033554455April 2022

0xea020004 = 3925999620

Nominal Operating Voltages and Currents

Ch.

Line

V

I

200AVDD6.0022.304
201DVDD5.9980.828

Performance Measures

Experiment=detdaq18, Run=754 (May 13, 2021, after removal from MEC....needs repair)

Summary

For Pedestals, method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as xxxx.



Mean Pedestal Value (ADU) Number of Ped NCPPedestal GraphsNoise (ADU)Number of Noise NCPNoise Graphs

4410 ± 36114

Histogram/Pixel Map

Pedestal NCP Map

34780

Histogram/Pixel Map

Noise NCP Map

Thermal measurements (all thermometers)

Configuration Settings

History

Run #

Experiment

Hutch

Shifts

Date Deployed





Testing 

Where

Who

Why

When (mm/dd/yy)

Notes
TIDO'Grady Timing tests12/08/21
TIDDoeringTest LCLS-II firmware12/06/21Dionisio couldn't write to pixel matrix
  • No labels