Identification Details and Hardware Changelog
Firmware | CarrierID0 | CarrierID1 | DigitalCardID0 | DigitalCardID1 | AnalogCardID0 | AnalogCardID1 | Date |
3925999620 | 0996432897 | 3590324234 | 1232100352 | 1154532875 | 2654088449 | 0033554455 | April 2022 |
0xea020004 = 3925999620
Nominal Operating Voltages and Currents
Ch. | Line | V | I |
---|---|---|---|
200 | AVDD | 6.002 | 2.304 |
201 | DVDD | 5.998 | 0.828 |
Performance Measures
Experiment=detdaq18, Run=754 (May 13, 2021, after removal from MEC....needs repair)
Summary
For Pedestals, method: Dark environment, nominal operation, collect 1000 frames @120Hz. Generate per-pixel mean and standard deviation over 1000 frames. Pixels more than 3 sigma out of detector-wide mean are flagged bad. Files archived as xxxx.
Mean Pedestal Value (ADU) | Number of Ped NCP | Pedestal Graphs | Noise (ADU) | Number of Noise NCP | Noise Graphs | |
---|---|---|---|---|---|---|
4410 ± 361 | 14 | 3 | 4780 |
Thermal measurements (all thermometers)
Configuration Settings
History
Run #
Experiment | Hutch | Shifts | Date Deployed |
---|---|---|---|
Testing
Where | Who | Why | When (mm/dd/yy) | Notes |
---|---|---|---|---|
TID | O'Grady | Timing tests | 12/08/21 | |
TID | Doering | Test LCLS-II firmware | 12/06/21 | Dionisio couldn't write to pixel matrix |