Performance Measures

Pedestal Summary

Method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as xxxx.

Map of pedestal non-conforming pixels for all gain modes here

Gain Mode

Mean Pedestal Value (ADU)

Number of NCP

Pixels Map, Histogram
FH2777 ± 124 29Graphs
FM2746 ± 12229Graphs
FL2732 ± 11930Graphs
AHL2782 ± 12329Graphs
AML2748 ± 12129Graphs


Noise Summary

Map of noise non-conforming pixels for all gain modes here.  

Gain ModeVariation (ADU)Number of NCPPixels Map, Histogram
FH5122Graphs
FM2680Graphs
FL25920Graphs
AHL6117Graphs
AML2283Graphs


Gain and uniformity (photon source)

Map of gain non-conforming pixels for all gain modes here

Gain ModeNumber of NCP
FH0
FM0
FL48
AHL10
AML6

Linearity (signal injection)

Thermal measurements (all thermometers)



  • No labels