Performance Measures

Pedestal Summary

Method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as xxxx.

Map of pedestal non-conforming pixels for all gain modes here

Gain Mode

Mean Pedestal Value (ADU)

Number of NCP

Pixels Map, Histogram
FH3012 ± 137 10Graphs
FM3012 ± 13710Graphs
FL3012 ± 13710Graphs
AHL3012 ± 13710Graphs
AML3013 ± 13710Graphs


Noise Summary

Map of noise non-conforming pixels for all gain modes here.  

Gain ModeVariation (ADU)Number of NCPPixels Map, Histogram
FH35Graphs
FM39855Graphs
FL3133812Graphs
AHL35Graphs
AML3729Graphs     


Gain and uniformity (photon source)

Map of gain non-conforming pixels for all gain modes here

Gain ModeNumber of NCP
FH6
FM6
FL135162
AHL6
AML6

Linearity (signal injection)

Thermal measurements (all thermometers)