The transportation locks are made of SST.


For motion test purposes:

  • In Air Motion Test:
    • Stages can be actuated in air
    • If motion test is done before bakeout then transportation locks need to be removed.
    • The screws that hold the transportation locks in place needs to be screwed back in before any motion test is performed.
    • These screws act as end stops. No motion test should be performed without the screws in place.
  • Bakeout:
    • Since the transportation locks are already taken out, no need to assemble those back on to the stages during bakeout.
    • Might be a good idea to place it in the same vacuum chamber during bakeout.
  • In Vacuum Motion Control Test:
    • SmarAct stages should not be energized during pump down or repressurizing to atmosphere. This was suggested by SmarAct.
    • SmarAct stages should only be energized at UHV vacuum pressure.
    • Motion test can be performed when vacuum level is UHV. No transportation lock is in place

  • No labels