Charge injection data
mfxc00118 r0172 epix10ka2m.1 MfxEndstation.0:Epix10ka2M.0
./calib/Epix10ka2M::CalibV1/MfxEndstation.0:Epix10ka2M.0/pedestals/172-end.data
./calib/Epix10ka2M::CalibV1/MfxEndstation.0:Epix10ka2M.0/pixel_gain/172-end.data
Decomposition of constants to image
Calibration constants of shape (7,16,352,384) were split for seven arrays (16,352,384)
- epix10ka2m_split_calib_for_7nda
Then, each of them was plot using
- geometry_image -g geometry-epix10ka2m.txt -a nda-16x352x384/nda-16x352x384-mfxc00118-r0172-epix10ka2m.1-pedestals-FH.txt -R0 -n3 -p5
Pedestals and gain for gain modes
FH
FM
FL
AHL-H
AHL-L
AML-M
AML-L
Summary
- Minor detector damage is observed in a few constant types - in charge injection gains for FL, AHL-L, AML-L
- Charge injection gains in stead of uniform constant gains can be used to eliminate this effect.