intro: this doc will walk you through starting up the RGA, gathering data, and sending it off for qualification. It assumes you are using an MKS brand RGA.



Qualification: Always go through Peter Stefan.

He is the official arbiter of chamber qualification, and he must have the final say. 

You must send Peter the RGA data file (on electron multiplier mode and accuracy 6) along with a CCG pressure reading (taken during data collection). 

If you do not do this, the chamber is not considered qualified from an administrative controls standpoint, and you will have bad luck for seven years. 



highest pressure for RGA use: 7.6E-6 torr

[or, if you want to use faraday mode (but why would you want to do that?), max pressure = 5 x 10-4 Torr]


connection issues? see Computer Recognition of RGA


Gathering data:

double check:

    • pressure is below 7e-10 torr
    • valve to RGA is open

heck, double check all valves!

software settings:

filament on

accuracy 6

electron multiplier

(w/ the lowest option in the drop-down menu)

saving data

to prevent the computer falling asleep: open an image (any image) and put it on slideshow

*insert video of RGA startup*


Email data to Peter Stefan (CC Brady Robertson and Cynthia Melendrez) for official qualification judgement:

  • The pressure that the cold cathode gauge reads as data collection stops. 
  • The data file.  Found here:
  • Cart number


documents:

  1. LCLS Procedure for Vacuum Qualification RGA Scans
  2. MKS RGA hardware manual

  3. MKS RGA software manual

  4. Amanda's doc: RGA Cart Maintenance and Operation