Here is detailed information for the experimental set-up:

Known Problems:

XTCS Newport Stages

URS150 rotational stage reference but will stall on move.  The UTS150 linear stage now works using "default" settings. A lot of time was spent between Christian, Eric, and Nick understanding connection and pin layout flipping throughout the cable routing from controller to stage.

The URS150 rotational stage will be replaced.

XTCS Filter Wheel

Update 30 July 2022: This is working. The following information is being for reference:

A new air-side DB15 to MForce cable was fabricated and installed today. Instead of a new IOC this device was connected to the user dumb motor patch panel and set up using the "XUV Filter Position" configuration.

The comments on the XTCS cable list for the filter wheel are:

  • Beckhoff, not a regular dumbmotor; probably go slightly longer (5-10 cm) but not any shorter. - same as the XUV spectrometer. (So we are using the correct parameters).

The device connects but no motion occurs at the stage.

The XTCS Filter Wheel configuration is:

  • MEC:USR:MMS:20 on user motor patch panel (Rack PR60)
  • Connections from outer flange to patch panel include custom cable and a MForce extension cable. (one flip)
  • Connections from inner flange to bulkhead are in-place DB15 (second flip) and a DB15 extension cable. 
  • Using XUV Filter position parameters

Internal XTCS cable pin-out:

Air-side cable pin-out:

MCC Side pin-out from MCC Schematic on the IMS and MForce Motors page:

The remaining ECS Delivery items (if not listed above) are:

Detectors:

  1. replacement for GIGE4
    1. replaced with GIGE8, known on camViewer and MEC screens as the beamline camera XRay_Imaging.
    2. needs mounting.

Motors:

  1. XTCS
    1. Newport rotational stage

Items originally listed but will be needed later for X523

Note: this was my understanding from Eric earlier this week, also see comment below on 25 July. 

  • e- spectrometer Opal 4K
  • thomson parabola Opal 4K

Items originally listed but no longer be used

  • XRD ePix10K
  • alignment camera 1
  • alignment camera 2
  • alignment probe


  • No labels

2 Comments

  1. Peregrine McGehee The timetool Opal 4K is needed for X455. Otherwise the rest is correct.

  2. Eric M Galtier Corrected. Also created new IOC for the XTCS filter wheel and added to User Dumb Motor screen.