Goal of this procedure: Enter oven mode and reach desired oven profile

Instructions: To keep track of changes to the procedure - copy this page, date it, and add execution notes in red. Remember to save changes.

Table of Contents:



                    NOTE - WORK IN PROGRESS!

Procedure

Checkout and setup


Procedure Execution notes
1Record goal density and helium buffer pressure

2

Set 10 Torr gauge set points to plus and minus 1 Torr of desired pressure

    • Note: the hysteresis value is where the trip occurs, the lower value is where the fault will clear
 
3Set the 1000 Torr gauge setpoint to 20 Torr

4

Confirm helium gas bottle is connected to fill line 1, regulator set to 5-10 psig 
5If using DPS, confirm it is operating nominally.

Procedure for oven turn on


Procedure Execution notes
1Verify vacuum conditions approximately 1e-6 Torr or better - Record pressure

2

Zero the 10 and 1000 Torr gauges

 

Purge fill line by running a small amount of He through the fill valve and then stopping and pumping out the IP - Record final pressure
3

Set EPS EXPERT Mode to be “ON” by turning the interlock key in the EPS box in Rack 20-27 in the gallery.

    • Note: EXPERT mode is “OFF” if the LED light is OFF and EXPERT mode is “ON” if the red light is ON)

4Unlock oven gate valves (switch to CAMAC mode) 
5Open oven gate valves 3183 and 3187. (there might be a small amount of helium between the valves). Continue pump down to 1e-6 Torr - Record pressure
6

Close angle valve VV3201 isolating the plasma beamline from the plasma turbo


7Check Beryllium window valve VV3208 is closed- close if not!
8

Set up strip-tool of pressure in the plasma oven:

    • VGCM:LI20:M3201:PMONRAW
    • VGCM:LI20:M3203:PMONRAW

9Fill IP with He to goal IP pressure. If using DPS then follow: How to perform a static fill with DPS

Hit EPS Reset to clear the low pressure fault

Set EPS EXPERT Mode to be “OFF” by turning the interlock key in the EPS box in Rack 20-27 in the gallery. (Note: EXPERT mode is “OFF” if the LED light is OFF and EXPERT mode is “ON” if the red light is ON) . 



While at the racks, Turn valve control for the beryllium window valve to local only and remove key. Note: the EPICS controls for this valve are also password protected so this will be very well configuration controlled.

Back at the computer, perform an EPS reset (this is needed to enable the heater)

Set up striptool of oven thermocouples plus the thermocouples on the bypass line, cooling water jackets, base-plate and air.

  • Saved as Li_oven_startup.stp




















  • No labels