SC3 or SSC: Serial Sample Chamber
- Location of the refocused 1 micron beam
- Used as diagnostics for gas run
cxi_sc3_inline_opal
camera is used to check beam on SC3 YAG- SC3 YAG is on PI3 Stage Motors
Mirror on Ap Stand can move into the beam and couple UV laser out for diagnostics.
Two important positions:
- Clear YAG
- Ap Stand = -9
- PI3 Y = 48.0326 (as of 2022-06-11, this position can change if changed YAG paddle)
- PI3 X = 10.8484 (as of 2022-06-11, this position can change if changed YAG paddle)
- When moving to this position, we should move the Ap Stand to the YAG position first, then insert the YAG to the down position
- Out-coupling Mirror for UV I0
- PI3 Y = 109 (to take YAG out)
- Ap Stand = -115
- When moving to this position, we should move YAG out first, then move the Ap Stand to the mirror position