**** Note: This page will be updated as camera is integrated into DAQ and IOCs are created ****
This is a how-to for a single camera. Ultimately, there will be two CCDs at qRIXS.
Initializing and cooling down camera
In order to operate the CCD, it needs to be cooled down first.
(images of startup screen, where to find cryotel and archon screens, highlight parameters to change)
Note that everything here is "expert mode", users (and POCs for the most part) should not need to access all of this information. Controls group is working IOCs with interlocks to simplify CCD operation.
Running camera in DAQ
Note: make sure to "Stop" acquisition when changing settings, and "Start" after to apply changes. (image highlighting Acquire button)
Settings to change | 1d | 2d |
---|---|---|
Readout Mode | FVB | Image |
Bins | 1, 1 | 1, 1 |
Image Size | 4800, 1 | 4800, 1200 |
Fixed Rate | 100Hz | 1Hz |
(images of archon screens, highlight fixed rate/binning/readout mode/what else?)
DAQ/AMI: