Post any graphs/presentations, we can figure out how to organize as things get dumped in here. For now, make sure to include which beamtime data/runs you looked at, your name, data of analysis.
Single Photon Gain results
Regardless the sagging area
Here are some beamtime energy (Beam primary energies and fluorescent energies) linearity study below. (Zongde, Philip)
Investigation on sagging pixels
Pixel [53,1] is a normal pixel, pixel [53,358] is in sagging area.
we can see the fitting line (forced across origin) deviates a lot compare with the normal pixel.
Some studies with @philip_adam_hart on the sagging region (high #column ), we take the photon counts ratio between single-pixel-cluster and single & double-pixel-cluster (run29, -50V), it seems like we have less two-photon cluster in sagging area (top: Module1, bottom: Module2)
We also checked the date at -110V (run 148, primary beam energy 11kev )