You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 37 Next »

Toggle showing or hiding all code blocks

Is there an issue? Escalation Path

  • Issue concerns X-ray beam delivery: see here
  • Issue concerns ECS: see here

Initial Spatial overlap between LPL, VISAR and X-rays

  • This procedure assumes that
    • the beamline is aligned up to yag3
    • the slits are aligned to the beam
    • the yag at TCC has been prealigned with the help of Questar 2 as well (for maximum accuracy) and the pin (the pin is used to measure the LPL spot size and set the plan of interaction)
  • Go to the yag at TCC

  • Remove the hutch Be CRL
  • Close slits 4 to 50 mic

  • Set a cross on Questar 1 screen and write down the X and Y positions in the preset spreadsheet of the experiment
  • Close the pulse picker and reinsert the Be CRL

  • Move to the pinhole

  • Tweak the position of the pinhole until it is centered on the cross

  • Save the new pinhole position

  • Move back to the yag

  • Send the VISAR laser with event code 43 and enable the trigger
  • Confirm that the focus of the VISAR provides a sharp and round image on any VISAR cameras (Visar gige 1 and 2)
  • Tweak the focus of the visar position accordingly (50 mic steps is good enough)

  • Turn off the VISAR trigger
  • Prepare the system to check the front alignment of the drive lasers

  • Check spatial overlap of the drive beams ABEF and GHIJ successively by centering the scattered signal on the cross of Questar 1 screen

  • Record each positions of the VISAR and drive lenses in the experiment spreadsheet

Initial Temporal overlap between LPL and X-rays

  • go to Ti sample

  • remove the Be CRL
  • open slit4 to 400 mic

  • set full X-ray transmission

  • open the pulse picker

  • Open the TCC scope

  • Set the EVR of the scope to 10Hz

  • Set channel 2 to 2 or 5 mV/div
  • Confirm that the X-ray pulse is seen around 60 ns.
  • Average over 100 sweep and save the trace on M2 (from C2)
  • Set the sweep averaging back to 1
  • Close the pulse picker

  • Insert the filter in front of Questar 1

  • Set the laser pulse shape to the timing shape

  • Send the 10Hz LPL (enable the trigger of the ns slicer) on target at full energy

  • Set the channel 2 to 10 mV/div
  • Confirm you see a trace on C2
  • Average over 100 sweep and save the trace on M3 (from C2)
  • Set the sweep averaging back to 1
  • Take a screenshot and post it to the elog
  • Set the channel 2 to 1V/div
  • Set the scope to single shot trigger

  • Further fine timing can be done by using a LiF coated window
    • Move to the appropriate target
    • Confirm Be CRL are inserted in the beamline
    • Set the timing of the drive to 0

    • Confirm that the streak window have zero offset in the window size requested

    • Take a single refence only shot to observe the change in reflectivity at the arrival time of the X-rays

    • To double check, confirm timing at another step

    • Fine tune the timing is the X-ray trace is not where it is expected using the nstiming.mv function accordingly.

Check Spatial overlap between LPL, VISAR and X-rays on target

  • Start by moving to the desired target (refer to the python manual here for details on the arguments)

  • Continue by checking the alignment of the drive lasers

    • Set the system in a mode where the laser energy is minimum, the filter is removed from questar 1 and the trigger is enabled

    • Confirm you see a scattering signal close to the X-ray cross (342, 561)
    • Move the hexapod X axis to center the beams on the cross and produce a round scattering signal (steps of 50 mic is good)

  • Continue by checking the alignment of the VISAR system
    • Adjust the VISAR Z axis to make sure the image is round and sharp (could be set to the value by using the right equation as well). Steps of 50 mic is good.

    • Alternatively, move the Z value by exactly the thickness of the ablator + material of interest and add the change in position due to the visar window. To calculate the change use the following command

Take reference images with VISAR and/or X-rays

  • For 5 references with VISAR and X-rays (at 20% max)

Take a driven shot with the LPL

  • For a driven shot at full energy (lpl_ener=1.0), with X-rays arriving 5 ns later (timing=5.0e-9) at full intensity (xray_trans=1) use the following command


  • No labels