Changing the quad settings

facet home → Profile Monitor → LI20 → SPEC LINE GUI

  • Leave big "Dipole" switch in "Without" setting (dipole energy is not adjusted)
  • Set "Z Object" to the position where the beam should be imaged, e.g., the center of the PB
  • Set "Z Image" to the position of the screen, e.g., z=2015.26 for DTOTR
  • M12, M34: transfer-matrix elements (set both to 0 for re-imaging)
  • Energy: determines where the focus is
  • Press "Calculate and Trim"
  • ATTENTION: there is no possibility to calculate without actually changing the magnets. There will be no elog entry! (2023, Nov 17)

Scripts for dealing with the spectrometer

Assumptions made in this script:

  • Magnet length: 1m
    LEFF_QS0 = 1; % [m]
    LEFF_QS1 = 1; % [m]
    LEFF_QS2 = 1; % [m]
  • Quad positions:
    z_QS0 = 1996.98249; % [m], middle of quad
    z_QS1 = 1999.206615; % [m], middle of quad
    z_QS2 = 2001.431049; % [m], middle of quad
  • Other important positions:
    1993.27370    # [m] FILS
    1992.82000    # [m] PIC_CENT
    2015.62984    # [m] LFOV
    2017.52998    # [m] PRDMP