Installation instructions

Installation instructions for PCI geometry: These instruction align the MXI optical axis parallel to the X-ray beam axis taking the chamber offset angle into account.
1.Use banking pad and mount to MXI instrument using two screws. Make sure to push banking pad into the two surfaces of the bottom frame as shown in image. (Image shows the ‘Close banking pad’).
2.Install shoulder screw to location shown on the image (Image shows location for the ‘Close position. See next slide for other positions.)
3.Push the banking pad against the  shoulder screw in order to align the center of BE lenses with FEL beam.
4.Secure the two clamping pucks at bottom of MXI to table or use clamps on both sides (MXI is instable if clamped on only one side).