Data
Detector/app/epix10ka_calib_components -t102 -S
*** parser.parse_args: Namespace(amax=None, amin=None, cframe=0, detname='epix10k2M', events=3, evskip=0, expname='mfxc00118', grindex=None, loglev='INFO', run=191, saveimg=True, segindex=None, stepnum=None, tname='102')
exp=mfxc00118:run=191:smd
epix10k2M
/reg/d/psdm/mfx/mfxc00118/calib
Access to calibration raw data
from Detector.PyDataAccess import get_epix_config_object, get_epix_data_object
odata = get_epix_data_object(evt, det.source)
crows = odata.calibrationRows()
calibrows shape:(16, 4, 384) size:24576 dtype:uint16 [3169 2982 3159 2822 3093...]
raw shape:(16, 352, 384) size:2162688 dtype:uint16 [3651 3470 3463 3534 3398...]
combine these arrays together for panel images with x20 rows for each calibration row.
Raw data images
Segments of entire detector
Segments of quad 0
Segments of quad 1
Segments of quad 2
Segments of quad 3
Issue
- panels of quad 0 behave as expected
- panels of quads 1,2,3 - calibration rows for "dark" and "max code" swap data???