Current Status

The Good

PMON (not a user area item but included for completeness)

PMON checked out - PROF:LI20:45. Both motion and light works - controlled on profile monitor panel. The camera is PROF:LI20:45.

sYAG

sYAG checked out - PROF:LI20:2434. Both motion and light works - controlled on profile monitor panel. The camera is CAMR:LI20:100.

USTHz

USTHz checked out but there is an issue with not having mechanical limit switches. Soft limits need to be used instead and checked before moving. For the short term, best to ask an expert to verify it is set up properly (C. Clarke, J. Allen, S. Gessner). Negative motion is okay when watching with the camera but real caution needs to be used with positive motion. OTRS:LI20:3070. Ladder edm updated but not checked out! Camera is CAMR:LI20:106. LED light is not installed but overhead lights were left on. 

USOTR

USOTR checked out - OTRS:LI20:3158. Ladder edm linked to from Profile Monitor panel will work. Camera is CAMR:LI20:101. LED light is set up on the USOTR channel accessible on the profile monitor edm panel.

IPOTR1

IPOTR1 checked out - OTRS:LI20:3180. Profile Monitor panel needs to be updated. Camera is CAMR:LI20:102. LED light is set up on the IPOTR channel accessible on the profile monitor edm panel however it isn't illuminating the target very well and will be improved.

DSOTR

DSOTR checked out - OTRS:LI20:3206. Profile Monitor panel needs to be updated. Camera is CAMR:LI20:103. LED light is set up on the DSOTR channel accessible on the profile monitor edm panel.

WDSOTR

WDSOTR checked out - OTRS:LI20:3230. Profile Monitor panel needs to be updated. Camera is CAMR:LI20:104. LED light is set up on the IP2A channel accessible on the profile monitor edm panel.

DTOTR2

User control only please... This is on MC01 Ch7 (X) and Ch8 (Y). DTOTR_x: Position 0 is out and 98 is in. DTOTR_y: 108 is YAG. CAMR:LI20:105.

LFOV

User control only please... This is on MC01 Ch6. Position >100 is out and 0 is in.  CAMR:LI20:107. ND 2.00 filter is on a flipper controlled by the Beckhoff Digital IO, channel 21, output 1 (Binary 5V Out) -  APC:LI20:EX02:5VOUT_1. Value "ON" = filter in.

PRDMP (not a user area item but included for completeness)

This is a fixed phosphor screen, no mover. CAMR:LI20:108.

Oven table mover

Fully checked out. Left with the bypass line in the beam path.

The Bad and the Ugly (aka work in progress)

Toroids

Tunnel work complete- toroids are connected to long haul cables. Need to be brought online and calibrated. 

IP Wirescanner 1

Checked out motion controls. PMT3350 and PMT3360 appear offline but are connected to long-hauls and no signs of why there would be issues from the tunnel side. Work still needed here.

IP Wirescanner 3

Checked out motion controls but LVDT is not calibrated. PMT3350 and PMT3360 appear offline but are connected to long-hauls and no signs of why there would be issues from the tunnel side. Work still needed here. 

Note, images of the card can be found here: https://drive.google.com/file/d/1scVopR61fpoo_dhtIH5aFozV2wHd0yiM/view?usp=sharing

IPOTR2 and IPOTR1P

Motion control is all good. Cables are prepared for the cameras. Just don't have cameras on these yet as they are not needed and we don't want to irradiate cameras without reason. IPOTR2 is controlled on OTRS:LI20:3202 and IPOTR1P is on OTRS:LI20:3175. 

Alignment laser injection

Pneumatic actuator works on OTRS:LI20:3158:PNEUMATIC. Needs to have limit switches fixed though so this has been mechanically blocked out in the meantime. Laser set up has been completed but picomotor controllers are not in EPICS. 

IP Wirescanner 1

Checked out motion but LVDT needs to be set up. Small question over the controls channel to be resolved. PMT3350 and PMT3360 appear offline. Work still needed here.

IP Wirescanner 3

Not checked out. (Note wirescanner #2 is not installed).

IPOTR2 and IPOTR1P

Motion control is all good. Just don't have cameras on these yet. IPOTR2 is controlled on OTRS:LI20:3202 and IPOTR1P is on OTRS:LI20:3175. Cameras currently not needed.

Alignment laser injection

Pneumatic actuator works on OTRS:LI20:3158:PNEUMATIC. Needs to have limit switches fixed though.