Discussion

georgi: 100eV photon scan range over 100 seconds. 1eV/s. few microeV change per shot at 1MHz. need precision 100 microev? need to take out shot intensity (normalization)

Alex R. says encoder has 100 microev resolution, scan even faster than 1eV/s required.

Corresponds to 10kHz measurements.

silke says that old lcls1 can't be reused. old usdusb one: relative encoder (quadrature pulses) new one: absolute encoder

alex says: new encoder can only read out at 10kHz, maybe 20kHz.

zach says: add a second encoder? relative quadrature encoders faster, maybe a MHz?

silke says: "usb4" relative encoder reader can go up to 5MHz, but maybe not enough space in the vacuum environment. alex found another renishaw that goes up to 15MHz.

zach asks: why not put it on the external shaft? georgi replies not straightforward to use that as the position of the grating. zach/Alex say it could be calibrated

cpo says: interpolate? georgi says the turnaround point would break interpolation. cpo says: mechanical motor turnaround time either fast or slow on timescale of 10kHz sampling should work.  in particular, turnaround time feels likely much slower than 10kHz sampling period?

alex w.: could add time-staggered 10kHz encoders to artificially increase rate

georgi says: need to understand the upper limits. interpolation may be too short-sighted. fast encoder is best.

alex wallace renishaw: https://www.renishaw.com/en/tonic-uhv-incremental-encoder-system-with-rslm20-linear-scale--11155

ankush encoder document: https://docs.google.com/document/d/1Ho1IpT4903wkUoJMV06z5hou9to5CfAgtJBDPdBl6I4/edit?ts=600b806d

      → (Silke) this lists the USB version of the encoder we originally used the PCI version for. We now have a box as Ankush laid out, except that it 'splits' the signal, so has in&out for each encoder channel. This device its listed towards the end of Supported Control & DAQ Devices

Matt says we may have an existing board that could do the readout/timestamping

Summary of Options/Information

  • 10kHz with interpolation?
  • multiple encoders?
    • combine absolute/relative (may be complexity issues)
    • 2 staggered slower encoders
  • need fpga timestamping
  • 100 microev resolution requirement

  • Optionmax ratework to be donedrawbacks/other consideration
    Aabsolute encoder used for relevant stage 10-20(?) kHz

    controls: rework PLC to send data (current solution likely ~1kHz)

    DAQ: likely FPGA timestamping


    B

    absolute encoder used for relevant stage, 

    interleave with second absolute encoder

    20-40 kHzas above

    need to ensure both encoders read back equivalent data,

    intersperse jitter from both?, same latency

    Cadd a relative quadrature encoder on shaft5 or more MHzDAQ: need board to read signal & timestamp

    complicated and possible not very reproducible mapping of measured &

    relevant motion

    Dadd a relative quadrature encoder in vacuum5 or more MHz

    DAQ: need board to read signal & timestamp

    engineering: figure out how to add relative encoder


To do

  • fill out the table options and costs (alex and silke)
  • talk to mechanical engineers (Georgi)
  • meet in two weeks

Feb. 5, 2021

Alex R., Georgi D., Chris F., Jana T., Matt W., Zach L., Joe R., Chris O.

Delay-stage scan discussion:

  • It was decided that for now RIX will not do on-the-fly delay stage scans, instead using standard DAQ step scans of an electronic delay (which incur some dead time when the delay is changed)
  • Georgi reserves the right to revisit this decision in the future

MHz encoder discussion:

  • Feels like option D (see above) is doable according to conversation between Georgi and Danny Morton
    • they meets with Axilon next week to verify (manufacturer of mono assembly))
  • Use this as our current default solution, but can still change (final decision in weeks)
  • Matt W. and Ryan H. think the camlink converter box could be used for readout
    • quadrature encoder "frame format" is understood and relatively simple (absolute encoders are more difficult)
    • quadrature encoder readout can be triggered
    • might need either a new camlink converter board with new connectors or some "connector transition"
  • currently no use case for high rate absolute encoders