Background Info tab

 

InstructionsAction Fields
Select Lead Tester's NamePulldown menu of Names
Enter any other testers' namesA box people can type in and the text is saved as a string
Enter dateMaybe don't need this as an enter box, but display of date/time might be nice and needs to be written to CSV
Select CryomodulePulldown with CM01-35 and HL1 & 2
Select CavityPulldown menu 1-8
Check that you're allowed to turn on the cavity, the CM is at 2K, and that the radiation monitoring system is installedCheckbox: Yup, we're good to go

Magnet Test tab

If this cryomodule's magnets haven't been tested yet:

InstructionsAction Fields
Open striptools to monitor magnet temperatures and readbacksButtons to open striptools (needs to pass CM/Cavity from background tab) or plots on screen
Push button to open display for quad magnet controlSuper fancy button that takes CM info and passes it to open the e/pydm display or have scripts to do next few steps
Reset and turn on the power supply. Degauss the magnet by selecting "Degauss" under the Ready menu (horrible instructions, apologies). After the degauss has finished, set BDES to 8.5 and trim.Scriptify please, need script to check for successful reset/on, degauss and trim and alert the user if any fail
For the XCOR and YCOR, reset and turn on the power supplies. Set BDES to 8.5 and trim.script. Same caveats as above. So these will be push buttons to run the scripts?
Leave the magnets on for at least an hour, monitoring the temps occasionally while continuing the cavity commissioning. After the hour has passed, trim the magnets to 0, then turn off the power supplies.Script to 0, trim, & off the magnets.
Take creen shots of the temp & readback plots and post to the elogA button to do this?