- Recommended Improvements:
- Holes not on meter side should be moved up to a line with holes on back for easier mounting of PCB
- Right now centered, should be off-center towards the lid
- Stand offs on SA, ML, and 3D printed are wrong height and location isn't optimal
- PN: 2102-440-AL
- Better location: Plate screws in center, see photos
- 3D printed version
- Walls are twice as thick, can't use lock washers in cut-in corner, use double-nut and excess solder to keep electrical connection tight
- Shell of SHV-connector on SA-261-101-32-C00a needs to be tied to ground, use solder lug, its not on ML
- PN: 031-759
- Original design counted on case being conductive, 3D version is no
- Recommend excess soldering on all nuts even if they have a lock-washer to hold in place better
- 2400 (210V limit)
- BOM
- Electrical
- Mechanical
- 24XX (230V +/-10 limit)
- 2410 (1.1kV limit) – typo on drawing, not 11kV limit