Detector Number | Type | Function | In DAQ? | Notes | Cabled? | Running? | Comments |
---|---|---|---|---|---|---|---|
1 | ePix100a | single photon diagnostics | Yes | in-vac | Yes | Yes | etch 12; vertical holder Q0 fibers/PDU |
2 | ePix100a | single photon diagnostics | Yes | in-vac | Yes | Yes | etch 25; horizontal holder Q1 fibers/PDU |
3 | ePix100a | XRTS | Yes | in-vac | No | Yes | ePix0 Q2 fibers/PDU Recovering from water spill. |
4 | ePix100a | Kalpha User spectrometer | Yes | in-vac | Yes | Yes | Q3 fibers/PDU |
5 | Zyla | MXI | Yes | in-air | Yes | Yes | zyla_xray_east Restart procedure at Zyla and iStar confluence page. Host is daq-mec-cam02. |
6 | XTCAV | Yes | in-vac | Yes | Yes |
ePix100a SPC vertical → Q0 fiber & PDU S0
ePix100a SPC horizontal → Q1 fiber & PDU S1
ePix100a XRTS → Q2 fiber & PDU S2
ePix100a Kalpha → Q3 fiber & PDU S3
Update on Fiber Mapping: (see 7.11 Opal Camera Configuration for fiber patch panel information)
- Commenting out epix100a_2_XRTS and setting "-i 2" for epix100a_3_Kalpha worked (but keeping -P 0x40) worked, and I checked via image inspection that I saw two different detectors.
- The epix100a persisted all day across nearly complete fiber changes - so the DAQ config is the issue (and the IOC was ok).
- In detail: Dimitri and I got into the fiber routing behind the Quad connections in PR60 since we suspect cable problems coming into cam01. So now cam01 has both the direct fiber bundle and the Fiberstore QSFP-SR4-40G that was in one of the two pgp02 ports. We don't need the quads again until late Feb, so can sort this out post L10011.