is shown on the following plot:

Left 4 plots show the charge injection curves ADC vs DAC (pedestals are NOT subtracted) for 4 energy ranges (LEX8, LEX1, HEX8, HEX1) obtained from July LCI run, right 4 plots - the same curves obtained from September run. In July all four curves were linear until saturation, in September LEX8 and LEX1 curves are unchanged, while HEX8 and HEX1 curves show negative sign signal with ~50 times smaller amplitude. pedestals are unchanged and the ratio of HEX8/HEX1 (after pedestal subraction) is still correct (~9) - this confirms that track and hold stages work and the failure happened somewhere in preamp. The small negative signal we see is probably the result of parasitic coupling between preamp input and output. 

...