Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Hit the "Power" button to start cooling the detector down. The minimum output power allowed by this device is 60W, and that alone will bring the camera down to colder temperatures than we want. So do NOT increase power. And DO NOT press "ON" or the CryoTel will try to get to the target temperature using its PID.

Archon

Image Modified

Archon controls are on 2 screens currently. The "parent" screen (left) is accessed through the home screen (labeled "CCD"), and the "child" screen (right) is accessed by pressing "STA5200".

...