Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

DetectorDAQ aliasConnectionCabled?In DAQ?
epix10k-540kquad_0

Fiber:

  • B2/#1,#2 
  • PR60 E15 Middle #1 
  • R62 E33 Right #2
  • MTP to daq-mec-pgp02
    • rear
    • upper right side
    • left most fiber port
  • Power:
B3/
  • Quad 0
  • vac cable -> Flange B3, Q0 feedthru -> air cable "epix10k quad 0" -> MPOD quad outlet 0
    • Note: vac cable labeled "PDU Cable Epix10k Q0" on quad side but "epix10k quad2" on flange side.
  • Fiber: using "epix10k quad 0" vacuum and air fibers. Flange B2, feedthru pair 0 -> PR60 patch panel quad port 0
YesYes
epix10k-540kquad_2
  • Power:  TBD -> Flange B3, Q2 feedthru -> air cable "epix10k quad 2" -> MPOD quad outlet 2
  • Fiber: using "epix10k quad 2" vacuum and air fibers. Flange B2, feedthru pair 2 -> PR60 patch panel quad port 2
  • Note on vacuum power cables:
    • Using cable "PDU Cable epix10k Q1": Minor alarm on DVDD when ON, HV error when failed to configure in the DAQ.
    • There is a second cable from Phil "epix10k quad 0" that I have not tested. It is on the optical table.
Fiber only, no power.No
epix10k-540kquad_3
  • Power:  Quad 3 vac cable -> Flange B3, Q3 feedthru -> air cable "C01 April 2024" -> MPOD quad outlet 3
  • Fiber: using "epix10k quad 3" vacuum and air fibers. Flange B2, feedthru pair 3 -> PR60 patch panel quad port 3
YesYesepix10k-540kquad_2

Details worked out with Detectors Group.

Fiber:

  • B2/#7,#8
  • PR60 E15 Middle #4 
  • R62 E33 Right #2
  • MTP to daq-mec-pgp02
    • rear
    • upper right side
    • left most fiber port

Power: B3/Quad 3

epix10k-540kquad_3

Fiber:

  • B2/#7,#8
  • PR60 E15 Middle #4 
  • R62 E33 Right #2
  • MTP to daq-mec-pgp02
    • rear
    • upper right side
    • left most fiber port
Power: B3/Quad 3

ePix100

SW Epix100a

"epix100a_f"

  • Power: Flange B3, South feedthru -> power cable E1 -> MPOD outlet 1
  • Fiber: using "epix100a epix 1" vacuum and air fibers. Flange D2 (left) -> PR60 patch panel epix port 1 → MTP to daq-mec-cam01
  • MECHome: "Epix 1"
  • DAQ: "epix100a_1_bxrts"
YesYes

ePix100

NE Epix100a

"epix100a_b"

  • Power: Flange B3, North feedthru -> power cable E0 -> MPOD outlet 2
  • Fiber: using "epix100a epix 0" vacuum and air fibers. Flange D2 (upper) -> PR60 patch panel epix port 2 → MTP to daq-mec-cam01
  • MECHome: "Epix 2"
  • DAQ: "epix100a_2_fxrts"
YesYes
Andor Zylazyla

in-air

USB/Fiber extender to daq-mec-cam02

YesYes
VISAR 1visar_1Network connection with Windows PC ioc-mec-visar1Yes

Yes

Not tested right now since under regular use

VISAR_2visar_2Network connection with WIndows PC ioc-mec-visar2Yes

Yes

Not tested right now since under regular use

Diode/Wave8

PVs are: 

  •  MEC:XT2-1:BMMON:*
  •  MEC:XT2-2:BMMON:*

TBD

Yes

Yes: see

Jira
serverSLAC National Accelerator Laboratory
serverId1b8dc293-975d-3f2d-b988-18fd9aec1546
keyECS-5105
.


Timing diode/QADC134

PVs are:

  • MEC:QADC:01:CH{0,1,2,3}

B3

ioc-mec-qadc.

Timing fiber connected.

Diode signal into A1 with SMA/Male - SSMC/Female Coaxial Jumper Cable and VLM-83-2W-S+ limiter.

Yes


Yes

XRT spectrometerhxsssN/AN/A