This concludes this procedure


Short term LPL Turn on procedure-(critical)

Turn off YFE

Do the following on the terminal:

  • ssh mec-laser
  • mecpython
  • import meclas
  • meclas.YFE.On() Note: Wait several minutes until the E-drives have turned on before proceeding to the next step. The E-drive turn on is confirmed after a series of chimes are heard. If not in the hutch, then the e-drive turn on can be confirmed by opening scope Lecroy A and changing the state of Long Pulse Laser Operation to 10Hz enable. The 10Hz trace will be visible when the E-drives have turned on (provided SHUTTER 4 on the LSM is open).
  • meclas.LPL.SHG_opt()


This concludes this procedure



Short term LPL Turn off procedure (non-critical):

Turn off YFE

Do the following on the terminal:

...