PV NamePurposeSave/restoreArchive
$(P):Firmwareread firmware number

$(P):HeartBeatread heartbeat
10s
$(P):HVStatusread status of high voltage
10s
$(P):Tempread temperature of chassis
10s
$(P):HVMonitorhigh voltage reading (V)
10s
$(P):HVSetpthigh voltage setting (V)

yes

10s
$(P):RackPositionString to hold a description of the rack positionyes
$(P):nn:DoseRateread back of dose from GM tube nn (01-10) in mR/hr (close enough, it saturates about 20Rsaturation curve starts to roll over about 10R/hr - see curve plot below)
1s
$(P):nn:Positionstring to hold a description of the position of GM tube number nnyes
$(P):Sequencesequence to readout the PVs at 1 Hz. The SEQ record has a delay of 0.06 seconds between sending 1 to each PV's PROC field so commands don't collide in the cable

...