Pretty much similar to epix10ka, pixek sizes 100x100um, 100x225um, 225x225um.

Geometry file

/cds/group/psdm/detector/alignment/epixhr2x2/geo-epixhr2x2-1-segment.data

Code Block
titleessential part of the geometry file
collapsetrue
# HDR PARENT IND        OBJECT IND     X0[um]   Y0[um]   Z0[um]   ROT-Z ROT-Y ROT-X     TILT-Z   TILT-Y   TILT-X

SEGMENT        0    EPIXHR2X2:V1  0         0        0        0       0     0     0    0.00000  0.00000  0.00000
IP             0    SEGMENT       0         0        0    10000      90     0     0    0.00000  0.00000  0.00000

Gain factors

Gain rangeDesign(*)Measured (***) Constants 1/g, ADU/eV

High124.4eV/ADU0.0410

Medium1/3
0.1230

Low1/80
3.279

*) Doering, Dionisio <ddoering@slac.stanford.eduSun 10/3/2021 7:39 PM

Hart, Philip Adam; Dubrovin, Mikhail; Blaj, Gabriel;Dragone, Angelo

O'Grady, Paul Christopher; Kamath, Umanath Ramachandra; Markovic, Bojan

Hi Philip, I just checked this with Angelo, the number we should use are the ones from the 10kT, which means:

H   M    L 
1  1/3   1/80

**) Dionisio: We will get the measured gain values with the  upcoming tests. For now we can estimate them using the expected gain ratios between high to medium and high to low gains respectively  x3.925 and x141.6.

***) Dionisio: Last time we measured we got for the Cd sealed source ~22000/900ADU = 24.4eV/ADU

Philip: I think 24 ev/ADU should be for high gain - 100 counts would be 2.4 keV, more or less the photon it's aimed at.