Slides from Dionisio Doering
Info from Matt
Weaver, Matt, Wed 9/22/2021 5:02 PM
Dubrovin, Mikhail, O'Grady, Paul Christopher
Hi Mikhail,
I just want to confirm what we talked about earlier today. The asicPixelConfig map which appears in the epixhr2x2 config object will be an array of shape (4,144,192). The ASICS are positioned like so:
A1 | A3
-----+------
A0 | A2
The map will be stored such that the orientation of all the ASICs appear the same.
-Matt
Sensor geometry
Pretty much similar to epix10ka, pixek sizes 100x100um, 100x225um, 225x225um.
Geometry file
/cds/group/psdm/detector/alignment/epixhr2x2/geo-epixhr2x2-1-segment.data
Gain factors
Gain range | Design(*), ~ADU/eV | Measured (***) , eV/ADU | Constants, ADU/keV | epix10ka default, ADU/keV | |
---|---|---|---|---|---|
High | 1 | 24.4eV/ADU | 41.0 | 16.40 | |
Medium | 1/3 | 13.7 | 5.466 | ||
Low | 1/80 | 0.512 | 0.164 |
*) Doering, Dionisio <ddoering@slac.stanford.edu> Sun 10/3/2021 7:39 PM
Hart, Philip Adam; Dubrovin, Mikhail; Blaj, Gabriel;Dragone, Angelo
O'Grady, Paul Christopher; Kamath, Umanath Ramachandra; Markovic, Bojan
Hi Philip, I just checked this with Angelo, the number we should use are the ones from the 10kT, which means:
H M L
1 1/3 1/80
**) Dionisio: We will get the measured gain values with the upcoming tests. For now we can estimate them using the expected gain ratios between high to medium and high to low gains respectively x3.925 and x141.6.
***) Dionisio: Last time we measured we got for the Cd sealed source ~22000/900ADU = 24.4eV/ADU
Philip: I think 24 ev/ADU should be for high gain - 100 counts would be 2.4 keV, more or less the photon it's aimed at.