camonitor ORTS:DMPH:695:RAW:ArrayData -#6
<hopefully data.....>
Some things to check if the Acqiris is causing problems:
13_11:46:45:2015-Nov-13 11:46:45,[STATUS ],[STATUS_DATA],-79.95,-79.35,-79.95,-79.75,-156.15,52.25,-860.95,-870.25,-774.25,33.05,49.85,49.85,276,273,275,36,0,0,1200,00:00:04
12_15:30:30:2015-Nov-12 15:30:30,[VERBOSE ],RxDetector::FrameAcquisitionFrameProcessor(83, 0x7f55d403bc20,0x7f55d42d2230) - starting
12_15:30:30:FrameReady: Frame #84 size=7372800 timestamp=59989 ms
13_11:38:45:2015-Nov-13 11:38:45,[STATUS ],[STATUS_DATA],-79.95,-79.35,-79.95,-79.65,-109.25,-109.95,34.05,42.65,45.45,34.75,49.85,49.85,276,273,275,37,0,0,1200,20:50:31
13_11:39:45:2015-Nov-13 11:39:45,[STATUS ],[STATUS_DATA],-79.95,-79.35,-79.95,-79.65,-109.25,-109.95,34.05,42.35,45.45,34.85,49.85,49.85,276,273,275,36,0,0,1200,20:51:31
13_11:40:45:2015-Nov-13 11:40:45,[STATUS ],[STATUS_DATA],-79.95,-79.35,-79.95,-79.75,-156.15,52.05,-860.85,-870.25,-774.15,34.75,49.85,49.85,276,273,275,37,0,0,1200,00:00:04
13_11:41:45:2015-Nov-13 11:41:45,[STATUS ],[STATUS_DATA],-79.95,-79.35,-79.85,-79.65,-156.15,52.05,-860.85,-870.25,-774.15,33.95,49.85,49.85,276,273,275,36,0,0,1200,00:00:04
To recover from that, you need to stop the DAQ process (ctrl+x in the telnet localhost 30099 terminal window) and open the Rayonix software capxure.
Reboot the detector controller (there is a button for that). Re-enable cooling. You might have to quite this process and start again to see the now reasonable temperatures. Once this looks fine, you can quit this software and restart the DAQ process: (ctrl+R in the telnet localhost 30099 terminal window)
...