Page History
...
epix10ka sensor central region between 4 ASICs
The internal gap between four ASICs in sensor,
...
- High (FH and AHL): 132 ADU/9.5 keV
- Medium (FM and AML): 43 ADU/9.5 keV
- Low (FL, AHL, AML): 1.32 ADU/9.5 keV
Gain factors
...
from charge injection default and measured
gain | charge injection | current default | measured (ADU / 9.5keV) |
---|---|---|---|
L | 0.46 | 0.01 | 1.32 |
M | 15. | 0.03(3) | 43 |
H | 46.7 | 1 | 132 |
References
...