Per-pixel gains from flat field measurement

DateDetectorMetrologyGain modePhoton sourcePhoton energy (keV)Occupancy (photons/pixel)Per-pixel histogram fit HDF5Deployed gain file
2016-02-16DsaCsPadhighIron K-alpha (6.4keV)7.20.015

/reg/g/psdm/detector/alignment/cspad/

calib-cxi-camera2-2016-02-05/calib/

CsPad::CalibV1/CxiDs2.0:Cspad.0/pixel_gain/

 run16_DsaCsPad_flat_orig.h5

cxi01516-r0016-2016-02-18-FeKalpha.data
2016-04-13DsaCsPadhighCu K-alpha (8.0keV)0.043

/reg/g/psdm/detector/alignment/cspad/

calib-cxi-camera2-2016-02-05/calib/

CsPad::CalibV1/CxiDs2.0:Cspad.0/pixel_gain/

 run6_DsaCsPad_flatIso.h5

run6_DsaCsPad_flatIso.data         

Note: CsPad has high and log gain modes. A factor of 7 is used to recalibrate the high gain to low gain, e.g. 28 ADU at high gain results in 4 ADU at low gain.

...