Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Trending calibGenCal outputs

Here is the trendiing trending of calibGenCal outputs for the 3 CU CAL modules. Three time phases are included: NRL (pre-CU) and Pisa (2 phases). The 3 files are FM109, FM119and FM101. Asymmetry amplitudes are VERY stable while slight variations are observed for pedestal position (~5 to 10 up to ~10 ADC units, ~5 ADC units in average) and MeV/DAC (~2% up to ~4%, ~1.5% in average), due to the change of power supply.

Trending CAL CPTs

Here is the trendiing of the 3 CU CAL module performances from CPTs runs taken at Pisa (2 suites) and at CERN (July, 7) : T01-FM109PS and SPS):
The 3 files are FM109-Pisa-PS-SPS, T02-FM119-Pisa-PS-SPSand T03-FM101-Pisa-PS-SPS for all time phases, and FM109-PS-SPS, T02-FM119-PS-SPSand T03-FM101 : no evolution was found-PS-SPS for the CERN phases only. Performances are found to be VERY stable at CERN, with slight differences wrt Pisa phases (pedestal position variation of ~5 ADC units, again due to change

 

August 7: Update of CU06 calibrations  pdf resumé

...