Page History
Issue mfxc00318 runs 10 and 13:
Panel 6 sick asic :
detdaq18 r92 offset_AHL and AML
mfxc00118 r92 offset_AHL and AML
Charge injection re-calibration using mfxc00118 runs 80
Code Block |
---|
Charge injection run 80
epix10ka_test_calibcycles exp=mfxc00118:run=80:smd --detname MfxEndstation.0:Epix10ka2M.0
Dark runs 82,86
epix10ka_test_calibcycles exp=mfxc00118:run=86:smd --detname MfxEndstation.0:Epix10ka2M.0
rm panels/*/*/*mfxc00118_r0080*
bsub -o 2020-08-14-r80-offset-log00 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i0
bsub -o 2020-08-14-r80-offset-log01 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i1
bsub -o 2020-08-14-r80-offset-log02 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i2
bsub -o 2020-08-14-r80-offset-log03 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i3
in ~3 hours
bsub -o 2020-08-14-r80-offset-log04 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i4
bsub -o 2020-08-14-r80-offset-log05 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i5
bsub -o 2020-08-14-r80-offset-log06 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i6
bsub -o 2020-08-14-r80-offset-log07 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i7
in ~3 hours
bsub -o 2020-08-14-r80-offset-log08 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i8
bsub -o 2020-08-14-r80-offset-log09 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i9
bsub -o 2020-08-14-r80-offset-log10 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i10
bsub -o 2020-08-14-r80-offset-log11 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i11
in ~3 hours
bsub -o 2020-08-14-r80-offset-log12 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i12
bsub -o 2020-08-14-r80-offset-log13 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i13
bsub -o 2020-08-14-r80-offset-log14 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i14
bsub -o 2020-08-14-r80-offset-log15 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i15
386 2020-08-14-r80-offset-log00
386 2020-08-14-r80-offset-log01
386 2020-08-14-r80-offset-log02
386 2020-08-14-r80-offset-log03
386 2020-08-14-r80-offset-log04
386 2020-08-14-r80-offset-log05
5250 2020-08-14-r80-offset-log06
386 2020-08-14-r80-offset-log07
386 2020-08-14-r80-offset-log08
386 2020-08-14-r80-offset-log09
386 2020-08-14-r80-offset-log10
386 2020-08-14-r80-offset-log11
386 2020-08-14-r80-offset-log12
386 2020-08-14-r80-offset-log13
386 2020-08-14-r80-offset-log14
386 2020-08-14-r80-offset-log15
In 2020-08-14-r80-offset-log06:
[W] L0249: pulser saw edges are not found, skip processing for ix41-iy24:
[W] L0249: pulser saw edges are not found, skip processing for ix42-iy24:
[W] L0249: pulser saw edges are not found, skip processing for ix43-iy24:
[W] L0249: pulser saw edges are not found, skip processing for ix44-iy24:
[W] L0249: pulser saw edges are not found, skip processing for ix45-iy24:
[W] L0249: pulser saw edges are not found, skip processing for ix46-iy24:
/reg/g/psdm/detector/gains/epix10k/panels/0000000001-0175696641-1056964630-1010951301-0019435010-0000000000-0000000000/offset/epix10ka_0032_20200808074716_mfxc00118_r0080_offset_AML.dat
Replace sick-asic with 0-s by mean values for 3 healthy-asics in
epix10ka_0032_20200808074716_mfxc00118_r0080_offset_AHL.dat
epix10ka_0032_20200808074716_mfxc00118_r0080_offset_AML.dat |
mfxc00118 80 offsets AHL, AML
mfxc00118 80 pedestals AHL-H, AML-M, AHL-L, AML-L
mfxc00118 80 zero-offsets in sick asic manually corrected to mean value for 3 other asics for AHL and AML
Pedestal re-calibration for mfxc00118 run 86
pedestals AHL-H, AML-M, AHL-L, AML-L
All gains - Gabriel's constants
This page is a folder for confluence documents about detector corrections. Use confluence left side window to navigate between documents.
Overview
Content Tools