chi2 for

...

AHL-H, AHL-L, AML-M, AML-L

panel 2

panel 2 tail of large values of chi2

Image AddedImage AddedImage AddedImage Added

panel 6

Image AddedImage AddedImage AddedImage Added

panel 6 tail of large values of chi2

Image AddedImage AddedImage AddedImage Added

Gain

Note

All single-panel gain is presented in raw units from the fit ADU/<event-number>. In combined  for detector plots gain is derived from calibration constants normalized on Gabriel's gain factors in ADU/keV (factor of 3 smaller than raw).


gain and negative tail for AML-L

panel 2

raw gain from fit in ADU/event

Image AddedImage Added

panel 6

raw gain from fit in ADU/event

Image AddedImage Added

WARNING: panel 6 one ASIC gain is almost twice smaller than ion other 3 ASICS ... we replace it by constant 0.164 ADU/keV, see Gain factors from Gabriel

gain for

...

AHL-H, AHL-L, AML-M, AML-L

panel 6

raw gain from fit in ADU/event

Image AddedImage AddedImage AddedImage Added

panel 2

raw gain from fit in ADU/event

Image AddedImage AddedImage AddedImage Added

other panels vertically jammed on a single plot 

normalized gain in ADU/keV for AHL-H, AHL-L, AML-M, AML-L and AML-L negative tail

Image AddedImage AddedImage AddedImage AddedImage Added

significant difference in ASIC gains is observed on panels 6 and 12

panel 12

AHL-L and AML-L

...

raw gain from fit in ADU/event


Image AddedImage AddedImage RemovedImage RemovedImage RemovedImage Removed