Identification
Firmware | xxxx | updated date | |
---|---|---|---|
Module identifiers | |||
Nominal Operating Voltages and Currents
Performance Measures
Bad Switch Status
...
Jungfrau bad gain mode switch status
exp:run |
---|
...
(date) | Number of bad pixels (total) | Number of bad switch pixels |
---|---|---|
cxix53120 : 35 (03/20/2022) | 4213, map | 2066, map |
cxilx5920 : 159 (03/07/2022) | 4564, map | 2067, map |
cxilw6120 : 207 (02/20/2022) | 4499, map | 1955, map |
cxic00120 : 42 (02/11/2022) | 3782, map | 1814, map |
cxic00120 : 2 (01/11/2022) | 857, map | 664, map |
exp=cxilv1019:run=142 (Nov 7, 2021)
mode is Normal
found 4193737 pixels in Normal
found 17 pixels in Med
found 550 pixels in Low
mode is Med
found 28 pixels in Normal
found 4194274 pixels in Med
found 2 pixels in Low
mode is Low
found 22 pixels in Normal
found 3 pixels in Med
found 4194273 pixels in Low
...
Pedestal
Method: Dark environment, nominal operation, collect 1000 frames @120Hz. Generate per-pixel mean and standard deviation over 1000 frames. Pixels more than 3 sigma out of detector-wide mean are flagged bad. Files archived as xxxxx.
Summary Values
Date | Mean | StdDev | #Bad |
---|---|---|---|
Pedestal Map
Noise map
Gain and uniformity (photon source)L
Linearity (signal injection)
Thermal measurements (all thermometers)
Experiment History
Experiment | Hutch | Shifts | Date Deployed |
---|---|---|---|
Commissioning | MFX | 1 | 15 June 2020 |
Calibration Records
Date | Hutch | Date | Record link | Comment |
---|---|---|---|---|
Commissioning | MFX | 7/29/20 | 15 June 2020 |