Switching gain modeAHL-HAHL-LAML-MAML-L
pedestal H(M)-from dark, L-evaluated3332182532952601
offset (B) from fit4584307736792985
offset (B) difference H(M)-L1507694
gain (G) from fit470.5215.40.49

Denote

  • p - pedestal
  • B - fit base-level offset parameter, <intensityADU> = B + G * <event-index-from-0-pulser>
  • G - gain

Current (plots above and results in table) naive version of L-pedestals evaluation

  • pL = pH - (BH-BL)

Possible correction for difference between B and pGabriel's version

  • pL = pH - (BH-BL) + BL - (BH-pH)*GL/GH,
    last term accounts for the difference between B and p in H- and L-gain modes, where
    • p - pedestal
    • B - fit base-level offset parameter
    • G - gain
  • this correction increase pedestals ~0.5-0.7%

Gabriel's version


Image Added

L-pedestals using Gabriel's formulae for AHL-L and AML-L

Image AddedImage AddedImage Removed

Questions

  • do we need to account for the difference between offset (B) and pedestal (Gabriel's version)
  • the same difference may be between fit B(L) and pedestal(L) OR it is smaller due to the gain? 

  • is current procedure correct?
  • if not
    • what is correct procedure?
    • or what could we calibrate with charge injection?

...

Panel 14(next-to-last) charge injection in ASIC1 does not work

2020-06-24 calibration constants for ALL gain ranges and panels 

detdaq18 run 52

pL were evaluated using Gabriel's formulae.

plots for pedestals, pixel_rms, pixel_status, pixel_gain (default), gain from charge injection

Image AddedImage AddedImage Added

Image AddedImage Added

deployed as

/reg/d/psdm/det/detdaq18/calib/Epix10ka2M::CalibV1/DetLab.0:Epix10ka2M.0/*/52-end.data

SHAPE  (7,16,352,384) : 7-gains, 16-panels of shape (352,384)