Bring up the CryoTel and Archon screens under the qRIXS tab in the RIX home screen.

Image Modified

CryoTel: 

Image RemovedImage Added 

Hit the "Power" button to start cooling the detector down. The minimum output power allowed by this device is 60W, and that alone will bring the camera down to colder temperatures than we want. So do NOT increase power. And DO NOT press "ON" or the CryoTel will try to get to the target temperature using its PID.

Archon

Image RemovedImage Added

Archon controls are on 2 screens currently. The "parent" screen (left) is accessed through the home screen (labeled "CCD"), and the "child" screen (right) is accessed by pressing "STA5200".

...

Settings to change1d2d
Readout ModeFVBImage
Bins1, 11, 12
Image Sensor Size 4800, 14800, 1200
LineScan ModeEnableDisabled
Fixed Rate100Hz1Hz 
Synced Policy/StatusINTERNALINTERNALLineScan ModeEnableEnable

The first 3 4 parameters can be found under "Readout" and "Collect" in the archon child screen, and "Fixed Rate" is last 2 are on the parent screen in the "Timestamps" section

Note from Dan Damiani: "Synced mode won't work properly in batched mode (all the frames in the batch will get the same timestamp). INTERNAL mode needs to be used for batched mode. This script can do it in an automated way: /cds/home/d/ddamiani/sync_timing.sh QRIX:STA:CCD:01 RIX:CCD:TPR:01:CH01  

Run Dan's script when changing modes


NOTE: make sure to "Stop" acquisition when changing settings, and "Start" after to apply changes.Image Removed  Image Removed

Image AddedImage Added  

^ These are the configurations for 1d mode. Configuration for 2d can be found here.


In order to turn the camera off and warm it up:

  1. Turn the cryotel OFF.
  2. Turn archon power OFF.
  3. Set the Heater to 10C.
  4. After it's warmed up (~2hrs), turn Heater OFF.


DAQ/AMI: 




NOTE: everything in the above screens is "expert mode". Users (and POCs for the most part) should not need to access all of this information. Controls group is working IOCs with interlocks to simplify CCD operation. 

...